![Tektronix ths3000 series Скачать руководство пользователя страница 31](http://html1.mh-extra.com/html/tektronix/ths3000-series/ths3000-series_technical-reference_1079180031.webp)
Performance Veri
fi
cation
16.
Press F4 TRIGGER OPTIONS to open the Trigger Options menu.
17.
Choose Video on 1 and then press Enter.
18.
Choose Polarity: Negative and then press Enter.
19.
Choose NTSC (or the standard that you want to test), then press Enter.
20.
Use the arrow buttons to select line number 310 for PAL or SECAM, or 262 for NTSC.
21.
Check to see that the oscilloscope triggers on line number 310 for PAL or SECAM, or 262 for NTSC.
Probe Calibration Generator Test
1.
Connect the test setup as shown. (See Figure 10.)
Figure 10: Probe calibration connections
2.
Press the Ch 1 button and then F3 to open the PROBE ON 1 menu.
3.
Set the Probe Type to Voltage and the Attenuation to 10:1.
4.
Press F1 - PROBE CAL and then press F4 - YES.
5.
Using the probe compensation adjustment on the probe, adjust the square wave response to a pure square wave
(pulse top must be straight).
6.
Press F4 - CONTINUE to start the DC calibration that is performed automatically.
7.
Repeat the test for the remaining channels on the oscilloscope.
The Probe Calibration Generator performance is veri
fi
ed when all of the on-screen functions
fi
nish successfully.
End of performance veri
fi
cation procedures.