Performance Veri
fi
cation
Test Setup
All tests should be performed with the batteries installed and the oscilloscope powered by the AC adapter.
1.
Connect the AC adapter to the oscilloscope and turn on the oscilloscope.
2.
Turn on the generators and let the probe and test equipment warm up for 20 minutes.
3.
Make a copy of the test record to tabulate the test results. (See page 20,
Using the Menus to Set the Input Channel Characteristics
For each channel that you test, you must set some parameters through the instrument menu. To access the menus and
settings, do the following:
1.
Press the Ch 1 button once (for channel 1). The menu that appears on the bottom of the screen allows you to change
settings using the buttons below the display (referred to as F1 through F4, from left to right). From left to right on
this menu, you can set the:
input on/off
input coupling
probe type and attenuation
polarity and bandwidth
A menu item highlighted in white indicates the current setting. A menu item highlighted in black indicates the
selection cursor is on that item.
In menus that have multiple options, use the arrow buttons to navigate to your selection and then press the Enter
button to set that selection.
2.
Press the
button once to access the Glitch and Zoom on/off controls.
3.
Press the
button again to access the MEASURE and ACQUIRE OPTIONS controls. In this menu, press the
MEASURE button to display a menu that allows you to select the characteristics and units of measure for the signals that
are applied to all of the instrument inputs.
For example, for MEASUREMENT A, select on Ch1, V dc. Then, press the MEASURE A B C D button to toggle
to MEASUREMENT B and then select on Ch2, V ac.
Also from this menu, pressing the ACQUIRE OPTIONS button displays a menu to select averaging options and other
waveform characteristics.