Performance Veri
fi
cation
Peak Measurement Accuracy Test
1.
Connect the test setup as shown in the previous test. (See Figure 3 on page 6.)
2.
Set all channels on the oscilloscope to:
Coupling: AC
Probe Type: Voltage, Attenuation: 1:1
Input Options: Polarity: Normal, Bandwidth: 20 MHz
Measure: Measurement on, Peak, Peak Type: Peak-Peak
Acquire Options: Glitch: Off, Memory: Short, Average: Off, Waveform: Normal
Ground level centered vertically
Horizontal scale: 1 ms/div
3.
Set the oscilloscope to display Channel 1.
4.
Set the oscilloscope vertical sensitivity to 100 mV/div.
5.
Set the calibration generator to 212.13 mV (0.6 V pp).
6.
Enable the calibration generator output.
7.
Verify that the measured output voltage on the oscilloscope is 0.56 Vpp to 0.64 Vpp.
8.
Repeat steps 3 through 7 for the remaining channels on the oscilloscope. Trigger on the appropriate channel as needed.
9.
Set the generator output to 0 V and to standby.
10