Specifications – TekRanger 2 (serial nos. B052000 and above)
1–74
TFS3031 TekRanger/TekRanger 2 Service Manual
Table 1–90: Horizontal System Dead Zone (Standard 1310 nm Singlemode, Options 04 and 06)
Characteristic
Performance Requirement
Supplemental Information
Attenuation Dead Zone
Pulsewidth
2000 m
1000 m
500 m
200 m
100 m
50 m
20 m
10 m
5 m
1 m
typical
2300 m
1350 m
600 m
290 m
175 m
100 m
46 m
36 m
23 m
18 m
15 m
Attenuation dead zone is defined
as the distance between the initial
and final points, where the
backscatter level is disrupted
beyond ±0.5 dB due to
encountering a single reflective
event.
Based on reflective event with no
greater than -40 dB reflectance.
Based on 9
µ
m, 0.1 NA SM fiber.
Event Dead Zone
Pulsewidth
2000 m
1000 m
500 m
200 m
100 m
50 m
20 m
10 m
5 m
1 m
typical
2250 m
1200 m
700 m
250 m
115 m
65 m
30 m
29 m
6 m
8 m
5 m
Event dead zone is defined as the
distance between the initial and
final points, where the reflected
pulse amplitude is -3.0 dB down
from its peak value.
Based upon reflective event with
no greater than -40 dB
reflectance.
OneĆmeter spec based on testing
to justify that the end of a 5Ćm
jumper can be detected.
Содержание TekRanger 2 TFS3031
Страница 5: ......
Страница 27: ...Module Level Service 1 4 TFS3031 TekRanger TekRanger 2 Service Manual...
Страница 39: ...Battery Recharging and Replacement 1 16 TFS3031 TekRanger TekRanger 2 Service Manual...
Страница 83: ...Specifications TekRanger serial nos below B052000 1 60 TFS3031 TekRanger TekRanger 2 Service Manual...
Страница 129: ...Accessories and Options 1 106 TFS3031 TekRanger TekRanger 2 Service Manual...
Страница 133: ...Theory of Operation 2 4 TFS3031 TekRanger TekRanger 2 Service Manual...
Страница 265: ...Replaceable Parts 6 14 TFS3031 TekRanger TekRanger 2 Service Manual...
Страница 270: ......
Страница 271: ......