Performance Check – TekRanger 2 (serial nos. B052000 and above)
TFS3031 TekRanger/TekRanger 2 Service Manual
4–49
1. Clean the connector on the TFS3031 and the TFS3031 multimode dynamic-
range test fixture.
2. Connect the TFS3031 to the dynamic-range test fixture and turn on power to
the instrument.
3. Check that loss at the front-panel connection is minimized (0.75 dB or less).
a. Press the
Setup
softkey to display the Test Setup menu, and set the
range to 1 km and the pulse width to 5 m. Select 1300 wavelength, or
850 if option 01. Press
Exit
.
b. Press and hold the
START/STOP
button to start a real-time acquisition.
c.
If front-panel loss is greater than 0.75 dB, clean the fiber connector and
the laser output port.
4. Press the
Setup
softkey, then press
SELECT
if necessary to display the Test
Setup menu.
5. Select 1300nm wavelength, or 850 nm if option 01.
6. Select the following test setups on the TFS3031:
Fiber Scan: Manual
Test Range: 10 km
Pulse Width: 1 m
Averages: approx. 3 minutes (as close as possible without being less than
three minutes)
7. Press the
Exit
softkey.
1-Meter Pulse Width
Содержание TekRanger 2 TFS3031
Страница 5: ......
Страница 27: ...Module Level Service 1 4 TFS3031 TekRanger TekRanger 2 Service Manual...
Страница 39: ...Battery Recharging and Replacement 1 16 TFS3031 TekRanger TekRanger 2 Service Manual...
Страница 83: ...Specifications TekRanger serial nos below B052000 1 60 TFS3031 TekRanger TekRanger 2 Service Manual...
Страница 129: ...Accessories and Options 1 106 TFS3031 TekRanger TekRanger 2 Service Manual...
Страница 133: ...Theory of Operation 2 4 TFS3031 TekRanger TekRanger 2 Service Manual...
Страница 265: ...Replaceable Parts 6 14 TFS3031 TekRanger TekRanger 2 Service Manual...
Страница 270: ......
Страница 271: ......