Performance Check – TekRanger 2 (serial nos. B052000 and above)
TFS3031 TekRanger/TekRanger 2 Service Manual
4–17
Power/charger adapter
Fiber test fixture
4. Check that loss at the front-panel connection is minimized (0.75 dB or less).
a. Press the
Setup
softkey to display the Test Setup menu, and set the
range to 1 km and the pulse width to 5 m. Press
Exit
.
b. Press and hold the
START/STOP
button to start a real-time acquisition.
c.
If front-panel loss is greater than 0.75 dB, clean the fiber connector and
the laser output port.
Содержание TekRanger 2 TFS3031
Страница 5: ......
Страница 27: ...Module Level Service 1 4 TFS3031 TekRanger TekRanger 2 Service Manual...
Страница 39: ...Battery Recharging and Replacement 1 16 TFS3031 TekRanger TekRanger 2 Service Manual...
Страница 83: ...Specifications TekRanger serial nos below B052000 1 60 TFS3031 TekRanger TekRanger 2 Service Manual...
Страница 129: ...Accessories and Options 1 106 TFS3031 TekRanger TekRanger 2 Service Manual...
Страница 133: ...Theory of Operation 2 4 TFS3031 TekRanger TekRanger 2 Service Manual...
Страница 265: ...Replaceable Parts 6 14 TFS3031 TekRanger TekRanger 2 Service Manual...
Страница 270: ......
Страница 271: ......