Battery Recharging and Replacement
TFS3031 TekRanger/TekRanger 2 Service Manual
1–13
NiCad Battery Recharging Procedure
To recharge the NiCad battery, plug the power/charger adapter into the port on
the top of the TFS3031, and into an appropriate active power source (see Figure
1–2).
The battery recharges as long as the power/charger adapter remains connected to
the instrument and power source.
Power source
(AC or DC)
Power cord,
battery clips, or
cigarette lighter
adapter
Power/charger
adapter
TEK RANGER
LASER
BATTERY
HELP
ON/OFF
START
STOP
Power/charger adapter plug
Flashing red LED indicates low battery.
Flashing green LED indicates battery
being recharged.
Figure 1–2: NiCad Battery Recharging Setup
H
A warning message flashed on the screen indicates that about 30 minutes of
normal operating time remain.
H
A flashing red
BATTERY
LED on the front panel indicates that about 10
minutes of normal operating time remain. (If the
BATTERY
LED is flashing
red at power-on and the LCD screen remains blank, there is not enough
battery power to run the instrument.)
H
The
BATTERY
LED flashes green when the power/charger adapter is
connected and the battery is recharging.
Содержание TekRanger 2 TFS3031
Страница 5: ......
Страница 27: ...Module Level Service 1 4 TFS3031 TekRanger TekRanger 2 Service Manual...
Страница 39: ...Battery Recharging and Replacement 1 16 TFS3031 TekRanger TekRanger 2 Service Manual...
Страница 83: ...Specifications TekRanger serial nos below B052000 1 60 TFS3031 TekRanger TekRanger 2 Service Manual...
Страница 129: ...Accessories and Options 1 106 TFS3031 TekRanger TekRanger 2 Service Manual...
Страница 133: ...Theory of Operation 2 4 TFS3031 TekRanger TekRanger 2 Service Manual...
Страница 265: ...Replaceable Parts 6 14 TFS3031 TekRanger TekRanger 2 Service Manual...
Страница 270: ......
Страница 271: ......