Specifications – TekRanger (serial nos. below B052000)
1–40
TFS3031 TekRanger/TekRanger 2 Service Manual
Table 1–43: Vertical System Dynamic Range (Extended Range 1310 nm Singlemode, Option 08)
Characteristic
Performance Requirement
Supplemental Information
Dynamic Range
Pulsewidth
2000
1000 m
500 m
100 m
50 m
20 m
10 m
5 m
1 m
SNR=1
Typical
30 dB
32 dB
28 dB
25 dB
28 dB
19.5 dB
18 dB
15 dB
13.5 dB
12 dB
7.5 dB
Dynamic range SNR=1 is defined
as the difference between the
extrapolated backscatter level at
the start of the fiber and the RMS
noise level.
Based on high APD and amplifier
gains. Low APD gain reduces 4
dB. Low amplifier gain reduces 5
dB.
Measured using a 4 km test fiber
(Corning SMFĆ28, 8.5
m
m core,
0.12 NA) after 3 minutes of
averaging.
A 0.5 dB front panel loss is
assumed.
Temperature Coefficient of
Dynamic Range
-0.15 dB/
o
C for T
25
o
C
Содержание TekRanger 2 TFS3031
Страница 5: ......
Страница 27: ...Module Level Service 1 4 TFS3031 TekRanger TekRanger 2 Service Manual...
Страница 39: ...Battery Recharging and Replacement 1 16 TFS3031 TekRanger TekRanger 2 Service Manual...
Страница 83: ...Specifications TekRanger serial nos below B052000 1 60 TFS3031 TekRanger TekRanger 2 Service Manual...
Страница 129: ...Accessories and Options 1 106 TFS3031 TekRanger TekRanger 2 Service Manual...
Страница 133: ...Theory of Operation 2 4 TFS3031 TekRanger TekRanger 2 Service Manual...
Страница 265: ...Replaceable Parts 6 14 TFS3031 TekRanger TekRanger 2 Service Manual...
Страница 270: ......
Страница 271: ......