Performance Tests
TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications
1–79
Figure 1–37: Line count accuracy correct result waveform
H
Disconnect all test equipment (TSG 121) from the digitizing
oscilloscope.
H
Turn off cursors by pressing CURSOR, then the main-menu
Function button, and, finally, Off from the side menu.
7. Check the Sync Duty Cycle.
a. Set up digitizing oscilloscope for Sync Duty Cycle Test:
H
Press TRIGGER MENU.
H
Press the Standard pop-up to select FlexFmt. Trigger Type should
already be set to Video.
H
Press the main-menu Setup.
H
Press the side-menu Field Rate.
H
Use the keypad to set the field rate to 60.05 Hz (press 60.05, then
ENTER).
H
Press the side-menu Lines.
H
Use the keypad to set the field rate to 793 lines (press 793, then
ENTER).
Содержание TDS 500C