Performance Tests
TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications
1–51
c.
Read results (Check against limits): See Figure 1–17.
H
CHECK that the Level readout in the main menu is within 9.863 V
to 10.137 V, inclusive, for the TDS 600B or is within 9.9393 V to
10.1147 V, inclusive, for the TDS 500C/700C.
H
Enter voltage on test record.
H
Press the main-menu button Slope; then press the side-menu button
for negative slope. See icon at left. Repeat substep b.
H
CHECK that the Level readout in the main menu is within 9.863 V
to 10.137 V, inclusive, for the TDS 600B or is within 9.9393 V to
10.1147 V, inclusive, for the TDS 500C/700C.
H
Enter voltage on test record.
Set DC calibration generator
to equal offset (+10 V).
2
Set vertical position to –3 divs.
Set vertical offset to +10 volts.
1
Read results.
3
Figure 1–17: Measurement of trigger-level accuracy
3. Confirm Delayed trigger system is within limits for Trigger-level/Threshold
accuracy:
a. Select the Delayed time base:
Содержание TDS 500C
Страница 4: ......
Страница 12: ...General Safety Summary viii TDS 500C TDS 600B TDS 700C Performance Verification and Specifications ...
Страница 15: ...Performance Verification Procedures ...
Страница 16: ......
Страница 20: ...Performance Verification Procedures 1 4 TDS 500C TDS 600B TDS 700C Performance Verification and Specifications ...
Страница 115: ...Specifications ...
Страница 116: ......
Страница 132: ...Nominal Traits 2 16 TDS 500C TDS 600B TDS 700C Performance Verification and Specifications ...
Страница 147: ......
Страница 148: ......