![Tektronix TDS 500C Скачать руководство пользователя страница 79](http://html1.mh-extra.com/html/tektronix/tds-500c/tds-500c_performance-verification-manual_1079136079.webp)
Performance Tests
TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications
1–63
c.
Modify the initialized front-panel control settings:
H
Set the horizontal SCALE to 200
s.
H
Press SET LEVEL TO 50%.
H
Use the vertical POSITION knob to center the display on screen.
H
Press SHIFT; then press ACQUIRE MENU.
H
Press the main-menu button Mode; then press the side-menu button
Average.
H
Select 128 averages with the keypad or the general purpose knob.
2. Confirm that the Probe Compensator signal is within limits for frequency:
a. Measure the frequency of the probe compensation signal:
H
Press MEASURE; then press the main-menu button Select
Measrmnt for Ch1.
H
Repeatedly press the side-menu button –more– until Frequency
appears in the side menu (its icon is shown at the left). Press the
side-menu button Frequency.
b. Check against limits:
H
CHECK that the CH 1 Freq readout is within 950 Hz to 1.050 kHz,
inclusive. See Figure 1–23.
H
Enter frequency on test record.
H
Press MEASURE; then press the main-menu button Remove
Measrmnt. Press the side-menu Measurement 1.
Содержание TDS 500C
Страница 4: ......
Страница 12: ...General Safety Summary viii TDS 500C TDS 600B TDS 700C Performance Verification and Specifications ...
Страница 15: ...Performance Verification Procedures ...
Страница 16: ......
Страница 20: ...Performance Verification Procedures 1 4 TDS 500C TDS 600B TDS 700C Performance Verification and Specifications ...
Страница 115: ...Specifications ...
Страница 116: ......
Страница 132: ...Nominal Traits 2 16 TDS 500C TDS 600B TDS 700C Performance Verification and Specifications ...
Страница 147: ......
Страница 148: ......