![Tektronix TDS 500C Скачать руководство пользователя страница 45](http://html1.mh-extra.com/html/tektronix/tds-500c/tds-500c_performance-verification-manual_1079136045.webp)
Performance Tests
TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications
1–29
Align the active cursor
over the waveform.
1
Read the
measurement
results.
2
Figure 1–4: Measurement of DC offset accuracy at zero setting
g. Test all channels: Repeat substeps a through f for all input channels.
3. Disconnect the hookup: No hookup was required.
WARNING.
The generator is capable of outputting dangerous voltages. Be sure to
set the DC calibration generator to 0 volts before connecting, disconnecting,
and/or moving the test hookup during the performance of this procedure.
Equipment
required
Two dual-banana connectors (Item 6)
One BNC T connector (Item 7)
One DC calibration generator (Item 9)
Two precision 50
W
coaxial cables (Item 5)
Prerequisites
The oscilloscope must meet the prerequisites listed on page 1–15
See Input Channels vs. Model on page 1–2.
Check DC Voltage
Measurement Accuracy
Содержание TDS 500C
Страница 4: ......
Страница 12: ...General Safety Summary viii TDS 500C TDS 600B TDS 700C Performance Verification and Specifications ...
Страница 15: ...Performance Verification Procedures ...
Страница 16: ......
Страница 20: ...Performance Verification Procedures 1 4 TDS 500C TDS 600B TDS 700C Performance Verification and Specifications ...
Страница 115: ...Specifications ...
Страница 116: ......
Страница 132: ...Nominal Traits 2 16 TDS 500C TDS 600B TDS 700C Performance Verification and Specifications ...
Страница 147: ......
Страница 148: ......