![Tektronix TDS 500C Скачать руководство пользователя страница 65](http://html1.mh-extra.com/html/tektronix/tds-500c/tds-500c_performance-verification-manual_1079136065.webp)
Performance Tests
TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications
1–49
H
CHECK that the Lower Limit readout, after the oscilloscope stops
triggering, is within 1.9
s to 2.1
s, inclusive.
H
Enter time on test record.
H
Use the keypad to return the Lower Limit to 500 ns and reestablish
triggering.
H
Press the side-menu button Upper Limit; then use the general
purpose knob to slowly decrease the Upper Limit readout until
triggering stops.
H
CHECK that the Upper Limit readout, after the oscilloscope loses
triggering, is within 1.9
s to 2.1
s, inclusive.
H
Enter time on test record.
4. Disconnect the hookup: Disconnect the cable from the generator output at
the input connector of CH 1.
Equipment
required
One DC calibration generator (Item 9)
One BNC T connector (Item 7)
Two 50
W,
precision coaxial cables (Item 5)
Prerequisites
See page 1–15.
DC Calibrator
50
W
coaxial cables
Dual banana to
BNC adapters
BNC T
connector
Digitizing Oscilloscope
Figure 1–16: Initial test hookup
1. Install the test hookup and preset the instrument controls:
a. Hook up the test-signal source:
H
Set the output of the DC calibration generator to 0 volts.
Check Accuracy,
Trigger-Level or
Threshold, DC Coupled
Содержание TDS 500C
Страница 4: ......
Страница 12: ...General Safety Summary viii TDS 500C TDS 600B TDS 700C Performance Verification and Specifications ...
Страница 15: ...Performance Verification Procedures ...
Страница 16: ......
Страница 20: ...Performance Verification Procedures 1 4 TDS 500C TDS 600B TDS 700C Performance Verification and Specifications ...
Страница 115: ...Specifications ...
Страница 116: ......
Страница 132: ...Nominal Traits 2 16 TDS 500C TDS 600B TDS 700C Performance Verification and Specifications ...
Страница 147: ......
Страница 148: ......