Performance Tests
TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications
1–17
Table 1–1: Test equipment (cont.)
Item number and
description
Purpose
Example
Minimum requirements
14. Generator, Video Signal
Provides PAL compatible outputs
Tektronix TSG 121
Used to Test Video Option 05
Equipped Instruments Only
15. Oscillator, Leveled Sine
wave Generator
60 Hz Sine wave
Wavetek 9100 with options
100 and 250 (or, optionally,
Tektronix SG 502)
Used to Test Video Option 05
Equipped Instruments Only
16. Pulse Generator
Tektronix CFG280
(or, optionally, PG 502)
Used to Test Video Option 05
Equipped Instruments Only
17. Cable, Coaxial
(two required)
75
, 36 in, male to male BNC
connectors
Tektronix part number
012-1338-00
Used to Test Video Option 05
Equipped Instruments Only
18. Terminator, 75
(two required)
Impedance 75
; connectors: female
BNC input, male BNC output
Tektronix part number
011-0102-01
Used to Test Video Option 05
Equipped Instruments Only
19. Generator, Sine Wave
100 kHz to at least 400 MHz. Variable
amplitude from 12 mV to 2 V
p-p
.
Frequency accuracy >2.0%
Rohde & Schwarz SMY
3
Checking Analog Bandwidth,
Trigger Sensitivity, Sample-
rate, External Clock, and
Delay-Time Accuracy
20. Meter, Level and Power
Sensor
Frequency range:10 MHz to 400MHz.
Amplitude range: 6 mVp-p to 2 V
p-p
Rohde & Schwarz URV 35,
with NRV-Z8 power sensor
3
Checking Analog Bandwidth
and Trigger Sensitivity
21. Splitter, Power
Frequency range: DC to 1 GHz.
Tracking: >2.0%
Rohde & Schwarz RVZ
3
Checking Analog Bandwidth
22. Generator, Function
Frequency range 5 MHz to 10 MHz.
Square wave transition time
25 ns.
Amplitude range: 0 to 10 V
p-p
into 50
Tektronix CFG280
Checking External Clock
23. Adapter (four required)
Male N to female BNC
Tektronix 103-0045-00
Checking Analog Bandwidth
24. Adapter
Female N to male BNC
Tektronix 103-0058-00
Checking Analog Bandwidth
25. Generator,
Leveled Sine Wave,
Medium-Frequency
(optional)
200 kHz to 250 MHz; Variable ampli-
tude from 5 mV to 4 V
p-p
into 50
Tegam/Tektronix SG 503
Leveled Sine Wave Genera-
tor
1, 3
Checking Trigger Sensitivity
at low frequencies
26. Generator, Leveled Sine
Wave, High-Frequency
(optional)
250 MHz to 1 GHz; Variable ampli-
tude from 5 mV to 4 V
p-p
into 50
;
6 MHz reference
Tegam/Tektronix SG 504
Leveled Sine Wave Genera-
tor
1
with SG 504 Output
Head
3
Checking Analog Bandwidth
and Trigger Sensitivity at high
frequencies
Содержание TDS 500C
Страница 4: ......
Страница 12: ...General Safety Summary viii TDS 500C TDS 600B TDS 700C Performance Verification and Specifications ...
Страница 15: ...Performance Verification Procedures ...
Страница 16: ......
Страница 20: ...Performance Verification Procedures 1 4 TDS 500C TDS 600B TDS 700C Performance Verification and Specifications ...
Страница 115: ...Specifications ...
Страница 116: ......
Страница 132: ...Nominal Traits 2 16 TDS 500C TDS 600B TDS 700C Performance Verification and Specifications ...
Страница 147: ......
Страница 148: ......