Performance Tests
TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications
1–65
H
Connect the Sense output of the generator through a second
dual-banana connector followed by a 50
W
precision coaxial cable to
the other side of the BNC T connector. Now connect the BNC T
connector to CH 1. See Figure 1–24.
DC Calibrator
50
coaxial cables
Dual banana to
BNC adapters
BNC T
connector
Digitizing Oscilloscope
Figure 1–24: Subsequent test hookup
e.
Measure amplitude of the probe compensation signal:
H
Press SHIFT; then press ACQUIRE MENU. Press the side-menu
button AVERAGE then enter 16 using the keypad or the general
purpose knob.
H
Adjust the output of the DC calibration generator until it precisely
overlaps the top (upper) level of the stored probe compensation
signal. (This value will be near 500 mV).
H
Record the setting of the DC generator.
H
Adjust the output of the DC calibration generator until it precisely
overlaps the base (lower) level of the stored probe compensation
signal. (This value will be near zero volts).
H
Record the setting of the DC generator.
f.
Press CLEAR MENU to remove the menus from the display. See
Figure 1–25.
Содержание TDS 500C
Страница 4: ......
Страница 12: ...General Safety Summary viii TDS 500C TDS 600B TDS 700C Performance Verification and Specifications ...
Страница 15: ...Performance Verification Procedures ...
Страница 16: ......
Страница 20: ...Performance Verification Procedures 1 4 TDS 500C TDS 600B TDS 700C Performance Verification and Specifications ...
Страница 115: ...Specifications ...
Страница 116: ......
Страница 132: ...Nominal Traits 2 16 TDS 500C TDS 600B TDS 700C Performance Verification and Specifications ...
Страница 147: ......
Страница 148: ......