Appendix A: MDO4000 Speci
fi
cations
Table 1: Analog channel input and vertical speci
fi
cations (cont.)
Characteristic
Description
The limits stated below are for ambient temperature of
≤
30 °C (86 °F) and the bandwidth selection
set to FULL. Reduce the upper bandwidth frequency by 1% for each °C above 30 °C.
Instrument
Volts/Div setting
Bandwidth
5 mV/div – 10 V/div
DC to 500 MHz
2 mV/div – 4.98 mV/div
DC to 350 MHz
MDO4104-3,
MDO4104-6
1 mV/div – 1.99 mV/div
DC to 175 MHz
5 mV/div – 10 V/div
DC to 500 MHz
2 mV/div – 4.98 mV/div
DC to 350 MHz
Analog
Bandwidth, 1 M
Ω
,
DC coupled
MDO4054-3,
MDO4054-6
1 mV/div – 1.99 mV/div
DC to 175 MHz
RMS noise (mV)
Instrument
Bandwidth selection
1 M
Ω
50
Ω
Full Bandwidth
≤
(300
μ
V + 8.0% of
Volts/div setting)
≤
(75
μ
V + 6.0% of
Volts/div setting)
250 MHz bandwidth
limit
≤
(100
μ
V + 5.0% of
Volts/div setting)
≤
(50
μ
V + 4.0% of
Volts/div setting)
MDO4104-3,
MDO4104-6
20 MHz bandwidth
limit
≤
(100
μ
V + 5.0% of
Volts/div setting)
≤
(50
μ
V + 4.0% of
Volts/div setting)
Full Bandwidth
≤
(130
μ
V + 8.0% of
Volts/div setting)
≤
(130
μ
V + 8.0% of
Volts/div setting)
250 MHz bandwidth
limit
≤
(100
μ
V + 6.0% of
Volts/div setting)
≤
(100
μ
V + 6.0% of
Volts/div setting)
Random Noise,
Sample Acquisition
Mode
MDO4054-3,
MDO4054-6
20 MHz bandwidth
limit
≤
(100
μ
V + 4.0% of
Volts/div setting)
≤
(100
μ
V + 4.0% of
Volts/div setting)
Table 2: Horizontal and acquisition system speci
fi
cations
Characteristic
Description
Long-term sample rate and
delay time accuracy
±5 ppm over any
≥
1 ms time interval
MDO4000 Series Oscilloscopes User Manual
181
Содержание MDO4054-3
Страница 2: ...x MDO4000 Series Mixed Domain Oscilloscopes ZZZ User Manual P071291300 071 2913 00...
Страница 3: ......
Страница 4: ...MDO4000 Series Mixed Domain Oscilloscopes ZZZ User Manual xx www tektronix com 071 2913 00...
Страница 11: ...Table of Contents iv MDO4000 Series Oscilloscopes User Manual...
Страница 19: ...Preface xii MDO4000 Series Oscilloscopes User Manual...
Страница 215: ...Appendix C P6616 General Purpose Logic Probe Information 196 MDO4000 Series Oscilloscopes User Manual...