Analyze Waveform or Trace Data
3.
Push the front-panel
Autoset
button.
4.
Push the front-panel
Test
button.
5.
Push the lower-bezel menu
Application
item. Turn multipurpose knob
a
to select
Limit/Mask Test
from the menu.
6.
Push the lower-bezel menu
Select Mask
item and, from the resulting side menu,
select
Limit Test
.
7.
Push
Create Limit Mask
from the
lower-bezel menu.
8.
On the resulting side-bezel menu, push
Source Channel
and turn multipurpose
knob
a
to choose the waveform to use as
the template for the limit test.
Source
Channel
a 1
9.
Push
Horizontal ±Limit
to set the mask
horizontal limits. The units are referenced to
graticule divisions, where one major division
contains 1,000 milli-divisions (mdiv).
Horizontal
±Limit
200mdiv
10.
Push
Vertical ±Limit
to set the mask
vertical limits. The units are referenced to
graticule divisions, where one major division
contains 1,000 millidivisions (mdov).
Vertical
±Limit
200mdiv
11.
Push
OK Create Limit Mask
to make the
mask in the oscilloscope.
OK
Create
Limit Mask
Select a standard mask.
1.
Push the front-panel
Test
button.
2.
Push the lower-bezel menu
Application
item. Turn multipurpose knob
a
to select
Limit/Mask Test
from the menu.
MDO4000 Series Oscilloscopes User Manual
155
Содержание MDO4054-3
Страница 2: ...x MDO4000 Series Mixed Domain Oscilloscopes ZZZ User Manual P071291300 071 2913 00...
Страница 3: ......
Страница 4: ...MDO4000 Series Mixed Domain Oscilloscopes ZZZ User Manual xx www tektronix com 071 2913 00...
Страница 11: ...Table of Contents iv MDO4000 Series Oscilloscopes User Manual...
Страница 19: ...Preface xii MDO4000 Series Oscilloscopes User Manual...
Страница 215: ...Appendix C P6616 General Purpose Logic Probe Information 196 MDO4000 Series Oscilloscopes User Manual...