Display Waveform or Trace Data
The trace options are:
1.
The Amplitude versus Time trace: The
instantaneous amplitude of the input,
after band-pass
fi
ltering, to the current
frequency range, as de
fi
ned by the
center frequency and span settings.
2.
The Frequency versus Time trace: The
instantaneous frequency of the input,
relative to the center frequency. The
vertical axis displays frequency, relative
to the center frequency.
3.
The Phase versus Time trace: The
instantaneous phase of the input, relative
to the center frequency. The vertical axis
is phase, with wrapping at r/-
180°.
All of these traces are derived from the
time-domain IQ data acquired for the RF
channel. They are time correlated with other
analog and digital channels and represent a
continuous time domain data stream.
The frequency versus time waveform handle
(baseline indicator) indicates the center
frequency. When the trace is above the
waveform handle, it is above the center
frequency. When the trace is below the
waveform handle, it is below the center
frequency.