Trigger Setup
Selecting Triggers
Trigger Type
Trigger Conditions
Edge
Trigger on a rising or falling edge, as de
fi
ned by the
slope control. Coupling choices are DC, LF Reject, HF
Reject, and Noise Reject.
Edge triggers are the simplest and most commonly
used trigger type, with both analog and digital signals.
An edge trigger event occurs when the trigger source
passes through a speci
fi
ed voltage level in the speci
fi
ed
direction.
Sequence
(B Trigger)
Combine an edge A Event (Main) trigger with the B
Event (Delayed) trigger to capture more complex signals.
(See page 91,
Using Sequence Trigger (A (Main) and B
Time.
After the A Event occurs, the trigger system waits
the speci
fi
ed amount of time, and then looks for the B
Event before triggering and displaying the waveform.
Events.
After the A Event occurs, the trigger system
looks for a speci
fi
ed number of B Events before
triggering and displaying the waveform.
Pulse Width
Trigger on pulses that are less than, greater than, equal
to, or not equal to a speci
fi
ed time. You can trigger on
positive or negative pulses. Pulse width triggers are
primarily used on digital signals.
Timeout
Trigger when no pulse is detected within a speci
fi
ed
time. The signal stays above or below (or either above
or below) a set value for a set amount of time.
Runt
Trigger on a pulse amplitude that crosses one threshold
but fails to cross a second threshold before recrossing
the
fi
rst. You can detect positive or negative (or either)
runts, or only those wider than, less than, greater than,
equal to, or not equal to a speci
fi
ed width. Runt triggers
are primarily used on digital signals.
84
MDO4000 Series Oscilloscopes User Manual
Содержание MDO4054-3
Страница 2: ...x MDO4000 Series Mixed Domain Oscilloscopes ZZZ User Manual P071291300 071 2913 00...
Страница 3: ......
Страница 4: ...MDO4000 Series Mixed Domain Oscilloscopes ZZZ User Manual xx www tektronix com 071 2913 00...
Страница 11: ...Table of Contents iv MDO4000 Series Oscilloscopes User Manual...
Страница 19: ...Preface xii MDO4000 Series Oscilloscopes User Manual...
Страница 215: ...Appendix C P6616 General Purpose Logic Probe Information 196 MDO4000 Series Oscilloscopes User Manual...