Analyze Waveform or Trace Data
Setting
Description
Select Action on Failure
Set how the oscilloscope responds to test failure. You can set multiple actions.
These are:
Stop acquisition
Save waveform to
fi
le
Save screen image to
fi
le
Print screen image
Aux out pulse
Set a remote interface service request (SRQ)
Select Action on Test Completion
Set how the oscilloscope will respond to test completion. You can set multiple
actions. They are:
Aux out pulse
Set a remote interface service request (SRQ)
Pre-Test Delay
Set a delay before starting a test.
Repeat Test
Set
On
for the test to repeat when it has run the minimum number of waveforms or
the minimum amount of time.
Set
Off
tor the test to run a single time and not repeat.
Mask Polarity
Set the mask polarity to use during the test. When
Both
is selected, the test will run
with
Normal
polarity for about half of the expected number of waveforms or amount
of time and then will run with
Inverted
polarity for the remained of the test.
Run the Test and View the Results
1.
Push the lower-bezel
Run Test
item to start
and stop the test.
Application
Limit/Mask
Test
Set Up
Mask
On
Select
Mask
Standard
Select
Standard
E1 Coax
Set Up
Test
1
Run Test
On|
Off
Show
Results
Off
2.
Push the lower-bezel
Show Results
item
and use the resulting side menu to select
whether to show basic or more detailed
results. You can also reset the results.
Quick Tips
Use Average acquisition mode to create a smoother, cleaner limit test mask.
If you want to re-use the mask later, save it to a
fi
le by selecting
Set Up Mask
from the lower menu and
Save Mask
to File
from the resulting side-bezel menu.
To simplify the setup of the test source, save the oscilloscope setup so that later you can quickly reload the settings to
correctly display the test source for limit testing.
Have the mask automatically re-scale with the source channel setting changes by selecting
Set Up Mask
from the lower
menu and
Lock to Source
as
On
with the resulting side-bezel menu.
The math waveform is not available when using mask testing.
MDO4000 Series Oscilloscopes User Manual
159
Содержание MDO4054-3
Страница 2: ...x MDO4000 Series Mixed Domain Oscilloscopes ZZZ User Manual P071291300 071 2913 00...
Страница 3: ......
Страница 4: ...MDO4000 Series Mixed Domain Oscilloscopes ZZZ User Manual xx www tektronix com 071 2913 00...
Страница 11: ...Table of Contents iv MDO4000 Series Oscilloscopes User Manual...
Страница 19: ...Preface xii MDO4000 Series Oscilloscopes User Manual...
Страница 215: ...Appendix C P6616 General Purpose Logic Probe Information 196 MDO4000 Series Oscilloscopes User Manual...