Analyze Waveform or Trace Data
1.
The zoomed-in portion of the analog
time is shown by brackets. In the
fi
gure
to the right, this time is outside of the RF
acquisition time.
2.
The RF acquisition time is shown by a
pair of vertical brackets.
3.
The spectrum time indicator turns gray
when there is no RF data to display, as
in the
fi
gure to the right. It will turn back
to orange if you move the spectrum time
so that it occurs inside the RF acquisition
time.
Limit and Mask Testing
Monitor an active input signal against a mask and output pass or fail results by judging whether the input signal is within the
bounds of the mask. To set up and run a limit or mask test, do the following:
1. Select or create the mask.
2. Set up the test.
3. Run the test and view the results.
Create or Select the Mask
The three mask types you can create or select are: limit test, standard, and custom.
Create a limit test mask.
1.
Push the front-panel
Default Setup
button.
2.
Connect your probe from your oscilloscope
to the mask source.
154
MDO4000 Series Oscilloscopes User Manual