Section 5: Command reference
Model 2461 Calibration and Adjustment Manual
5-8
2461-905-01 Rev. A / February 2016
:CALibration:LOCK
This command locks calibration constants.
Type
Affected by
Where saved
Default value
Command and query
Not applicable
Nonvolatile memory
0
Usage
CALibration:LOCK
CALibration:LOCK?
Query returns
0
= Calibration unlocked
1
= Calibration locked
Details
Use this command to prevent comprehensive adjustment.
Send the
CALibration:LOCK?
query to determine if the calibration constants are locked.
Calibration constants are locked if
1
is returned; calibration constants are unlocked if
0
is returned.
If you change the Model 2461 calibration password and forget it, you must return the Model 2461 to
the local Keithley Worldwide Service Center to be reset. There is no user-serviceable procedure to
restore the default calibration password on the Model 2461.
Example
:CAL:LOCK
Lock the calibration constants.
Also see
(on page 5-9)