Section 5: Command reference
Model 2461 Calibration and Adjustment Manual
5-10
2461-905-01 Rev. A / February 2016
:CALibration:SAVE
This command saves calibration constants to nonvolatile memory after completing the adjustment procedure.
Type
Affected by
Where saved
Default value
Command and query
Not applicable
Nonvolatile memory
Not applicable
Usage
:CALibration:SAVE
:CALibration:SAVE?
Details
Use this command to store internally calculated calibration constants that were derived during the
comprehensive adjustment procedure to nonvolatile memory. Calibration constants are retained
indefinitely once saved.
The
:CALibration:SAVE
command is usually sent after all other adjustment steps have been
completed, but before sending the
:CALibration:LOCK
command.
Adjustment is only temporary until the :
CALibration:SAVE
command is sent to permanently store
the calibration constants.
Adjustment data will not be saved if one of the following conditions occurs:
•
Calibration was not unlocked by sending the
:CALibration:UNLock
command before
adjustment
•
Invalid data exists (for example, an adjustment step failed or was aborted)
•
An incomplete number of adjustment steps were performed (for example, if a negative full-scale
step was omitted)
Example
:CAL:SAVE
Save the calibration constants.
Also see
(on page 5-8)