Model 2461 Calibration and Adjustment Manual
Section 5: Command reference
2461-905-01 Rev. A / February 2016
5-3
:CALibration:ADJust:SENSe
This command adjusts the sense function.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Nonvolatile memory
Not applicable
Usage
CALibration:ADJust:SENSe <range>
CALibration:ADJust:SENSe <negZero>
CALibration:ADJust:SENSe <negFullScale>
CALibration:ADJust:SENSe <posZero>
CALibration:ADJust:SENSe <posFullScale>
<range>
The sense range to adjust
<negZero>
The negative zero value of the selected range (from digital multimeter reading)
<negFullScale>
The negative full-scale value of the selected range (from digital multimeter reading)
<posZero>
The positive zero value of the selected range (from digital multimeter reading)
<posFullScale>
The positive full-scale value of the selected range (from digital multimeter reading)
Details
Calibration constants must be unlocked with the
:CALibration:UNLock
command before sending
this command.
To use this command:
1. Set the source output value.
2. Select the range to adjust.
3. Send the
CALibration:ADJust:SENSe
command with each of the parameters in the usage
table above.
4. Repeat this procedure for each of the ranges.
If the specified parameters do not fit within the allowed ranges for the parameter, error -222,
"Parameter data out of range" is generated. Once the Model 2461 has successfully selected the
appropriate parameter range, it checks to see if autorange for the active sense function is enabled. If
it is enabled, error -221, "Settings conflict" is generated. If no error occurs, that active sense function
range point is adjusted using the corresponding parameter.
You must send the
:CALibration:SAVE
command after making changes, or your changes will not
be saved through a power cycle.
Parameters for a given sense function and range may be sent in any order. However, once one
CALibration:ADJust:SENSe
command executes, the other parameters must also be executed or
an error -200, “Execution error” is generated when the
CALibration:SAVE
command is processed
at the end of the adjustment procedure.