Model 2461 Calibration and Adjustment Manual
Section 4: Adjustment
2461-905-01 Rev. A / February 2016
4-5
Current calibration adjustment data table
Source range
Source current
DMM current reading*
1 μA
+1.0000 μA
____________________________
μA
+0.0000 μA
____________________________
μA
−
1.0000 μA
____________________________
μA
−
0.0000 μA
____________________________
μA
10 μA
+10.000 μA
____________________________
μA
+00.000 μA
____________________________
μA
−
10.000 μA
____________________________
μA
−
00.000 μA
____________________________
μA
100 μA
+100.
00 μA
____________________________
μA
+000.00 μA
____________________________
μA
−
100.00 μA
____________________________
μA
−
000.00 μA
____________________________
μA
1 mA
+1.0000 mA
____________________________ mA
+0.0000 mA
____________________________ mA
−
1.0000 mA
____________________________ mA
−
0.0000 mA
____________________________ mA
10 mA
+10.000 mA
____________________________ mA
+00.000 mA
____________________________ mA
−
10.000 mA
____________________________ mA
−
00.000 mA
____________________________ mA
100 mA
+100.00 mA
____________________________ mA
+00.000 mA
____________________________ mA
−
10.000 mA
____________________________ mA
−
00.000 mA
____________________________ mA
1 A
+100.00 mA
____________________________ A
+000.00 mA
____________________________ A
−
100.00 mA
____________________________ A
−
000.00 mA
____________________________ A
* DMM reading used in corresponding adjustment step. See the adjustment procedure in
(on page 4-7).