In this section:
Introduction .............................................................................. 5-1
Model 2461 calibration verification and adjustment commands 5-1
Introduction
This section contains detailed information about the Model 2461 remote calibration verification and
adjustment commands, error messages, and methods to detect the end of each adjustment step.
See
(on page 4-1) for detailed calibration adjustment procedures.
Model 2461 calibration verification and adjustment commands
Model 2461 calibration verification and adjustment commands are used to save calibration constants,
lock-out calibration, and program date parameters.
:CALibration:ADJust:COUNt?
This command queries the number of times the Model 2461 has been adjusted.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
CALibration:ADJust:COUNt?
Details
Use this command to determine the total number of times the instrument has been adjusted.
Example
:CAL:ADJ:COUN?
Request the calibration adjustment count.
Output:
2
Indicates that the instrument has been adjusted
two times.
Also see
(on page 5-2)
Section 5
Command reference