Section 4: Adjustment
Model 2461 Calibration and Adjustment Manual
4-16
2461-905-01 Rev. A / February 2016
Current range calibration adjustment commands
Step
Command/procedure*
Description
1
:SOUR:CURR:RANG <Range>
Select the source range.
2
:SOUR:CURR -<Range>
Establish negative polarity.
3
DMM reading 1.**
Using the DMM, measure voltage across the shunt
resistor.
4
(calculated I) = V (DMM reading 1) / R
Calculate the shunt resistor current.***
5
:CAL:ADJ:SOUR <calculated I>
Adjust the source function negative full scale.
6
:SYSTem:ERRor?
Check Model 2461 for errors.
7
:CAL:ADJ:SENS <calculated I>
Adjust the sense function negative full scale.
8
:SYSTem:ERRor?
Check Model 2461 for errors.
9
:SOUR:CURR 0.0
Set the output to 0 A.
10
DMM reading 2.**
Using the DMM, measure the voltage across the
shunt resistor.
11
(calculated I) = V (DMM reading 2) / R
Calculate the shunt resistor current.
12
:CAL:ADJ:SOUR <calculated I>
Adjust the source function negative zero.
13
:SYSTem:ERRor?
Check Model 2461 for errors.
14
:CAL:ADJ:SENS <calculated I>
Adjust the sense function negative zero.
15
:SYSTem:ERRor?
Check Model 2461 for errors.
16
:SOUR:CURR +<Range>
Establish positive polarity.
17
DMM reading 3.**
Using the DMM, measure voltage across the shunt
resistor.
18
(calculated I) = V (DMM reading 3) / R
Calculate the shunt resistor current.
19
:CAL:ADJ:SOUR <calculated I>
Adjust the sense function positive full scale.
20
:SYSTem:ERRor?
Check Model 2461 for errors.
21
:CAL:ADJ:SENS <calculated I>
Adjust the source function positive full scale.
22
:SYSTem:ERRor?
Check Model 2461 for errors.
23
:SOUR:CURR 0.0
Set the output to 0 A.
24
DMM reading 4.**
Using the DMM, measure the voltage across the
shunt resistor.
25
(calculated I) = V (DMM reading 4) / R
Calculate the shunt resistor current.
26
:CAL:ADJ:SOUR <calculated I>
Adjust the source positive zero.
27
:SYSTem:ERRor?
Check Model 2461 for errors.
* Perform the complete procedure for each range, where <Range> = 4, 5, and 7. There are no
adjustment steps for the 10 A range.
** The
<DMM_Reading
n
>
parameter is the DMM reading from the previous step.
*** To get more accurate readings when using a shunt resistor, measure the resistance of the shunt resistor (rather
than using manufacturer specifications) and use that value in the shunt resistor current calculation.