Model 2461 Calibration and Adjustment Manual
Section 5: Command reference
2461-905-01 Rev. A / February 2016
5-11
:CALibration:UNLock
This command unlocks the calibration constants.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Nonvolatile memory
Not applicable
Usage
:CALibration:UNLock "<password>"
<password>
The calibration password; the default password is
KI002400
Details
Use this command to unlock the calibration constants so that comprehensive adjustment can be
done.
Send the
CALibration:LOCK?
query to determine if the calibration constants are locked.
Calibration constants are locked if
1
is returned; calibration constants are unlocked if
0
is returned.
The password parameter must be enclosed in quotation marks.
Example
:CAL:UNL "KI002400"
Unlock the calibration constants.
Also see
(on page 5-8)
(on page 5-9)