Model 2461 Calibration and Adjustment Manual
Section 5: Command reference
2461-905-01 Rev. A / February 2016
5-9
:CALibration:PASSword
This command changes the calibration password.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Nonvolatile memory
KI002400
Usage
CALibration:PASSword "<password>"
<password>
The calibration password; can contain up to 8 characters; default is
KI2400
Details
Use this command to change the calibration password, which is used to lock the calibration
constants.
To change the password, you must first send the existing password, and then send the command
again with the new password.
The password parameter must be enclosed in quotation marks.
Example
:CAL:PASS "KI002400"
:CAL:PASS "Cal123"
Send the existing password.
Send the command again with the new
password; password is changed to
Cal123
.
Also see
(on page 5-8)