Specifications
4–8
VX1410 IntelliFrame Instruction Manual
Table 4–6: Electromagnetic Compatibility (EMC)
1
Characteristic
Description
Emissions
Emissions shall be within the limits specified by the
following requirements:
Enclosure:
EN 55011 Class A limits for radiated emissions
AC Mains:
EN 60555-2 AC power harmonic emissions
EN 55011 Class A limits for conducted emissions
Immunity, Enclosure, Radio
Frequency Electromagnetic Field
No mainframe failures when the mainframe is subjected to
a 3 V/m electromagnetic field over the frequency range of
27 MHz to 500 MHz.
Immunity, Enclosure, Electrostatic
Discharge (ESD)
Up to 8 kV with no change to control settings or
impairment of normal operation, and up to 15 kV with no
loss of stored data or damage that prevents recovery of
normal operation by the user.
Immunity, Fast Transients, Com-
mon Mode
No loss of stored data, change to control settings,
degradation of performance, or temporary loss of function
will occur when the mainframe is subjected to the following
transients:
Port
Peak Voltage Tr/Th
Rep Frequency
Signal & Control
0.5 kV
5/50 ns 5 kHz
DC Power
0.5 kV
5/50 ns 5 kHz
AC Power
1.0 kV
5/50 ns 5 kHz
FCC
The mainframe complies with the radiated and conducted
emissions requirements of FCC 47 CFR, Part 15,
Subpart B, Class A equipment.
1
If the VX1410 IntelliFrame is configured with Option 1D, Delete Power Supply, all
product specifications in this table do not apply. To be VXI compliant, the power
supply must adhere to the specifications of the VXI standard.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com