Specifications
VX1410 IntelliFrame Instruction Manual
4–3
Table 4–2: Secondary Power
1, 2 (Cont.)
Characteristic
Description
Maximum DC Load Ripple/Noise
Voltage
Ripple/Noise
+24 V
150 mV
pp
+12 V
50 mV
pp
+5V
50 mV
pp
–2 V
50 mV
pp
–5.2 V
50 mV
pp
–12 V
50 mV
pp
–24 V
150 mV
pp
Maximum DC Induced Ripple/Noise
Voltage
Ripple/Noise
+24 V
150 mV
pp
+12 V
50 mV
pp
+5V
50 mV
pp
–2 V
50 mV
pp
–5.2 V
50 mV
pp
–12 V
50 mV
pp
–24 V
150 mV
pp
Protection
Over voltage protection
Under voltage protection
Over temperature protection
Short circuit protection
Voltage Readout
Readout of
±
24 V,
±
12 V, +5 V, –2 V, –5.2 V, +5 V,
External, and +5 V Standby if present through VXIbus or
RS-232 (with Option 1M)
Voltage Readout Accuracy, typical
±
3%
1
If the VX1410 IntelliFrame is configured with Option 1D, Delete Power Supply, all
product specifications in this table do not apply. To be VXI compliant, the power
supply must adhere to the specifications of the VXI standard.
2
If the VX1410 IntelliFrame is configured with Option 1D, Delete Power Supply, to
maintain safe operation, the power supply that provides power to the chassis should
not exceed the I
MP
current ratings found under DC Current Capacity (I
MP
) on
page 4–2. The short circuit capacity of each voltage should not exceed 120% of the
I
MP
rating. The total power dissipation of the chassis, with modules installed, should
not exceed 925 Watts.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com