background image

STATus Subsystem

3–66

VX1410 IntelliFrame Instruction Manual

STATus:QUEStionable:VOLTage:PTR
STATus:QUEStionable:VOLTage:PTR?

Sets or queries the power supply positive transition filter.

You can only set or clear bits 3 and 4 of the transition filter. Setting a bit enables
the corresponding power supply condition to propagate to the event system. Any
failures will display on the front panel readout. If you disable the bits, any failure
in the corresponding power supply will not be reported to the front panel readout
or to the event system.

Bits 0 – 2 and bits 5 – 8 of the transition filter are always enabled. Bit 15 always
returns a 0. The factory default setting is 32743. The +5 V Standby and +5 V
External are not monitored.

STATus:QUEStionable:VOLTage:PTR <Mask>

STATus:QUEStionable:VOLTage:PTR?

<Mask>

Query Response

{<NRf> | <Nondecimal Numeric>}

 N 

 #HFFFF

<NR1>

Voltage Positive Transition Filter Bits

Definition

D3

+5 V Standby

D4

+5 V External

Not Applicable

–222, “Data Out of Range”
You attempted to set the transition filter to an illegal value.

None

Command:

STAT:QUES:VOLT:PTR #H7FF7

Query:

STAT:QUES:VOLT:PTR?

Response:

32759

STATus:QUEStionable:VOLTage:CONDition?

Syntax

Parameters

Reset Value

Errors and Events

Dependencies

Examples

Related Commands

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Содержание IntelliFrame VX1410

Страница 1: ...ess underutilized and idle equipment along with credit for buybacks and trade ins Custom engineering so your equipment works exactly as you specify Critical and expedited services Leasing Rentals Demo...

Страница 2: ...instructions are for use by qualified personnel only To avoid personal injury do not perform any servicing unless you are qualified to do so Refer to the Safety Summary prior to performing service Ar...

Страница 3: ...es that in all previously published material Specifications and price change privileges reserved Printed in the U S A Tektronix Inc P O Box 1000 Wilsonville OR 97070 1000 TEKTRONIX and TEK are registe...

Страница 4: ...fect failure or damage caused by improper use or improper or inadequate maintenance and care Tektronix shall not be obligated to furnish service under this warranty a to repair damage resulting from a...

Страница 5: ...r RS 232 Connector 3 4 Backplane Connectors 3 5 Enhanced Monitor 3 9 Command Groups 3 13 Display Commands 3 13 Status Commands 3 13 System Commands 3 15 Test Commands 3 18 VXI Monitor Commands 3 18 IE...

Страница 6: ...7 2 Removal and Replacement 7 4 Troubleshooting Procedures 7 16 Repackaging Instructions 7 17 Theory of Operation 7 18 Replaceable Parts Parts Ordering Information 8 1 VX1410 IntelliFrame Common Repla...

Страница 7: ...talling the Empty Slot Panel Fillers 1 20 Figure 1 11 Locations of Safety and Chassis Ground Screws on Left Rear of the Mainframe 1 21 Figure 1 12 Installing the Backplane EMI DIN Shields 1 22 Figure...

Страница 8: ...r of the Mainframe 7 5 Figure 7 3 Removing the Fan Assembly 7 6 Figure 7 4 Removing the Fans 7 6 Figure 7 5 Removing the Monitor Board 7 7 Figure 7 6 Removing the Power Supply Interface Board 7 8 Figu...

Страница 9: ...5 105 and 135 155 Jumper Location on the Standard Monitor Board B 3 Figure B 3 Enhanced Monitor Board Jumpers B 4 Figure B 4 J20 and J21 Backplane Interface Connector B 6 Figure B 5 Power Supply Inter...

Страница 10: ...IEEE 488 2 Common Commands 3 19 Table 3 13 Parameter Types for Syntax Descriptions 3 23 Table 3 14 Default Factory RS 232 Communication Settings 3 77 Table 3 15 Effects of PRESet on Serial Port Param...

Страница 11: ...ise Check 5 2 Table 6 1 PC RS 232 Settings for the Calibration Procedure 6 2 Table 7 1 Fuses 7 16 Table 8 1 VX1410 IntelliFrame Common Replaceable Parts 8 2 Table C 1 SCPI Conformance Information C 1...

Страница 12: ...ensure that the product is properly grounded To avoid electric shock or fire hazard do not operate this product with covers or panels removed To avoid fire hazard use only the fuse type and rating sp...

Страница 13: ...cates an injury hazard immediately accessible as you read the marking WARNING indicates an injury hazard not immediately accessible as you read the marking CAUTION indicates a hazard to property inclu...

Страница 14: ...uscitation is present To avoid electric shock disconnect the main power by means of the power cord or if provided the power switch Dangerous voltages or currents may exist in this product Disconnect p...

Страница 15: ...traits and typical characteristics of the mainframe H Adjustment Procedures provide service related command information for use with the Enhanced Monitor option H Performance Verification includes the...

Страница 16: ...ower supply has a volt ampere capacity of 1660 VA but only 872 watts is usable by VXI instrument modules at any one time see Appendix A Power Budget Worksheet The mainframe accommodates exclusively di...

Страница 17: ...1410 IntelliFrame has an autoconfigurable solid state backplane that uses electronic jumpering of the IACK and BUS GRANT signal lines There are no jumpers to move or install on the backplane This resu...

Страница 18: ...a rear RS 232 interface The Option 1M also includes a 20 character alphanumeric display on the front panel with scroll keys for manual viewing for mainframe status Table 1 1 lists the VX1410 IntelliF...

Страница 19: ...plug type varies according to country see Table 1 2 H Line fuse 15 A 250 V H Line fuse 20 A 250 V H Line fuse 6 3 A 250 V H VXI Plug Play Floppy disk See the Readme file on disk for installation infor...

Страница 20: ...ale X DB9 female serial null modem Tektronix part number 012 1379 XX H AC to DC low voltage converter supply to externally supply power to the Enhanced Monitor Options 1M and 2M Tektronix part number...

Страница 21: ...Frame comes standard with two power cords and three fuses Before installing the mainframe you must determine the correct fuse and power cord for your application Use the following information to deter...

Страница 22: ...ule installation instructions The VX1410 IntelliFrame is designed to operate on a bench or in a rackmount environment For proper cooling allow at least two inches 5 1 cm of clearance on the rear and s...

Страница 23: ...ble tray depth of 1 75 inches Similarly you can install Option 2U to provide a cable tray depth of 3 5 inches Figure 1 3 on page 1 10 shows how to install Option 2U to provide a cable tray depth of 4...

Страница 24: ...all feet optional Install butch plate optional Install cable tray and tighten 4 screws Remove 4 inside screws 4 2 6 5 Figure 1 2 Installing the Option 1U 1 75 inch Deep Cable Tray Option 2U 3 5 Inch D...

Страница 25: ...feet Install butch plate optional Install cable tray and tighten 4 screws Install feet optional 4 3 2 Figure 1 3 Installing the Option 2U 4 25 inch Deep Cable Tray Option 1U 2 5 Inch Installation is...

Страница 26: ...ble tray and tighten 4 screws Install butch plate optional Install 4 screws do not tighten Install feet optional 1 Remove 4 feet 7 6 5 4 Figure 1 4 Installing the Option 2U 6 0 Inch Deep Cable Tray Op...

Страница 27: ...e recesses into the rack The steps for installing the rackmount kits consist of installing the rackmount hardware to the mainframe and then installing the mainframe in the rack You will need the follo...

Страница 28: ...surface Latch is located at the rear Bracket 2 Screws 5 Screws Bracket VXI Plug Play Position Key Handle 2 Screws Track Figure 1 5 Installing the Rackmount Kit Hardware 4 Remove the five 8 32 screws f...

Страница 29: ...pair of holes that mount the rear bracket so the flange to flange distance see Figure 1 6 matches the front rail to rear rail spacing that you just measured in step 2 NOTE If necessary refer to the r...

Страница 30: ...ance is only valid for the mainframe without the rubber feet and cable trays Refer to Figures 4 2 4 3 and 4 4 on pages 4 10 4 11 and 4 12 for information on vertical clearances of the mainframe with c...

Страница 31: ...ck assembly to the front and rear rails Seat the screws fully but lightly to adjust the mounting later 9 Tighten the screws left loose in step 4 to fix the front to rear flange spacing of the slide ou...

Страница 32: ...inet 15 Tighten the four screws that you left loose at the rear of the rack when you performed step 8 and pull the mainframe part way out of the rack Tighten the 10 32 screws using 28 inch lbs of torq...

Страница 33: ...inges Rackmount Door Rackmount Bracket 2 Flat Head Screws Catch 2 Pan Head Screws Rackmount Bracket 2 Flat Head Screws Figure 1 9 Installing the Rackmount Door Option 2R Artisan Technology Group Quali...

Страница 34: ...UL listed to UL 3111 1 and CAN CSA C22 2 No 1010 1 and is IEC 1010 1 compliant You should use VXIbus modules with the mainframe that are listed or designed to limit electric shock and fire H Investig...

Страница 35: ...vate effectively robbing airflow from installed modules Installing the fillers provides improved cooling for installed modules improved EMI shielding and more accurate air temperature sensing to contr...

Страница 36: ...the power supply to the mainframe You can use the chassis ground connection to connect the grounds of one or more instruments to the mainframe This ensures a common ground connection between instrumen...

Страница 37: ...you intend to install the Backplane EMI DIN shields 3 Refer to Figure 1 12 and install the shields over the backplane connector 4 Mount the connector in place using two 4 20 thread forming screws 5 R...

Страница 38: ...odules The shield conforms with the VXI specifica tion for intermodule shielding 7 Install a blank EMI front panel for each empty slot making sure that the EMI shielding is in contact with the adjacen...

Страница 39: ...position of FF This allows the mainframe to dynamically configure the address However make sure that no other module uses the same slot address as the enhanced monitor board determined by the positio...

Страница 40: ...clude the Optional Monitor you can measure the power supply voltages with a digital voltmeter Connect one lead of the voltmeter to the supply pins of the 25 pin connector on the rear panel of the main...

Страница 41: ...the mainframe Slot 0 is reserved for the mainframe controller all other slots are available for any VXIbus modules IntelliGuides on the bottom of each slot automatically open when you install a module...

Страница 42: ...display allow you to scroll through the display selections Power Switch Figure 2 2 Partial Front View of the VX1410 IntelliFrame with Option 1M Figure 2 3 shows the rear view of the standard VX1410 In...

Страница 43: ...WARNING Do not connect an RS 232 cable to the 25 pin connector The connector is not an RS 232 connector Connecting an RS 232 cable to the connector can result in damage to the equipment Enhanced Moni...

Страница 44: ...or message Once all error messages have been retrieved the READY indicator stops flashing and remains on The amber ACCESSED indicator lights when the MODID line is accessed by the slot 0 device Refer...

Страница 45: ...Sub D connector lets you monitor the power supply voltages fan speed and the maximum slot temperature rise within the mainframe The connector also provides remote on and off capability and access to t...

Страница 46: ...put signal proportional to the maximum temperature rise of the 13 modules 100 mV C 12 Tach1 A square wave output signal proportional to the speed of Fan 1 RPM 15 Pulse Width where Pulse Width is the p...

Страница 47: ...n 1M and 2M If you use pins 8 and 21 5 VB to supply 5 V Standby to the backplane make sure that the rear panel jumper is in the correct position refer to Enhanced Monitor Board Jumpers in Appendix B 4...

Страница 48: ...cludes a 9 pin RS 232 connector that allows connec tion to a RS 232 host Figure 3 2 shows the pinouts of the 9 pin RS 232 connector Table 3 2 describes the pin assignments Pin 9 Pin 1 Figure 3 2 RS 23...

Страница 49: ...Signal Ground GND 6 No Connection 7 Request to Send RTS 8 Clear to Send CTS 9 No Connection Backplane Connectors Table 3 3 shows the P1 connector pinouts for all slots in the VX1410 Intelli Frame Tab...

Страница 50: ...SFAIL 11 GND BG3OUT BERR 12 DS1 BR0 SYSRESET 13 DS0 BR1 LWORD 14 WRITE BR2 AM5 15 GND BR3 A23 16 DTACK AM0 A22 17 GND AM1 A21 18 AS AM2 A20 19 GND AM3 A19 20 IACK GND A18 21 IACKIN SERCLK A17 22 IACKO...

Страница 51: ...A31 LBUSC04 12 LBUSA05 GND LBUSC05 13 5 2 V 5 V 2 V 14 LBUSA06 D16 LBUSC06 15 LBUSA07 D17 LBUSC07 16 GND D18 GND 17 LBUSA08 D19 LBUSC08 18 LBUSA09 D20 LBUSC09 19 5 2 V D21 5 2 V 20 LBUSA10 D22 LBUSC10...

Страница 52: ...LBUSC04 12 MODID07 GND LBUSC05 13 5 2 V 5 V 2 V 14 MODID06 D16 LBUSC06 15 MODID05 D17 LBUSC07 16 GND D18 GND 17 MODID04 D19 LBUSC08 18 MODID03 D20 LBUSC09 19 5 2 V D21 5 2 V 20 MODID02 D22 LBUSC10 21...

Страница 53: ...stem status as determined by the monitoring system Use the NEXT and PREV buttons on the side of the display to scroll through the individual messages Table 3 6 lists some of the messages that you may...

Страница 54: ...s 24 Volts XX XXV 12 Volts XX XXV 5 Volts X XXV 5V Stby X XXV 5V External X XXV 2 Volts X XXV 5 2 Volts X XXV 12 Volts XX XXV 24 Volts XX XXV Temperature Status Messages Ambient T XX C S0 S1 XX XX C S...

Страница 55: ...t to the current interval You can also query the individual messages and status information using the commands and queries described in the remainder of this chapter The remainder of this chapter prov...

Страница 56: ...Commands Header Description Sets or queries user defined messages to display on the front panel readout Status Commands Commands in the STATus subsystem along with several IEEE 488 2 Common Commands...

Страница 57: ...ntents of the Questionable Condition register Sets or queries the contents of the Questionable Status Enable register Sets or queries the filter cleaning interval time Returns the remaining time for t...

Страница 58: ...ns the next event code from the Error Event queue Returns a comma separated list of all unread event codes from the Error Event queue Returns the number of unread events from the Error Event queue Set...

Страница 59: ...o default values Configures the RS 232 port parameters for use with a computer Configures the RS 232 port parameters for use with a terminal Sets or queries the number of stop bits sent with each char...

Страница 60: ...query returns whether the power is on or off Queries the number of times the mainframe power has been cycled Queries the power source of the enhanced monitor Sets or queries whether the VXI SYSRESET s...

Страница 61: ...failure code for the last self test command that was executed Returns a failure code as a string describing the last executed self test command and the test results VXI Monitor Commands Commands in t...

Страница 62: ...nable register Returns the contents of the Standard Event Status Register Returns the identification message of the mainframe Synchronizes command execution with the controller Sets or queries the pow...

Страница 63: ...The standard simplifies the task of programming a group of instruments that use SCPI Instead of having to learn different commands for every instrument the programmer may use the same commands for the...

Страница 64: ...Root node Leaf nodes Figure 3 3 Tree Hierarchy The header path and leaf node together form the command header Figure 3 4 shows the command header for the leaf node indicated in Figure 3 3 Root Node He...

Страница 65: ...ric Integers in hex H octal Q or binary B format command parameter H00FF B010101 Q0753 NR1 Integers response parameter 0 1 15 NR2 Decimal numbers response parameter 1 2 3 141516 6 5 NRf Integer decima...

Страница 66: ...ou can enter commands in either upper or lower case You can precede any command with white space characters White space characters include a space or any combination of the ASCII control characters he...

Страница 67: ...SYSTem VERSion TEST RESult If the command that follows the semicolon has the same header path you may omit the colon and the path and state only the new leaf node This makes it possible for example t...

Страница 68: ...to mainframes with Option 1M Please note that Option 2M does not include a display Use the DISPlay subsystem to display user defined messages on the front panel Figure 3 6 lists the DISPlay subsystem...

Страница 69: ...s Parameter Query Response String Block String Not Applicable 223 Too Much Data You attempted to send too much data to the display 220 Parameter Error You attempted to send nonprinting data in the blo...

Страница 70: ...STATus subsystem The STATus subsystem controls the parameters shown in Figure 3 7 Figure 3 7 STATus Subsystem Hierarchy Use the STATus subsystem to monitor the status of the voltages temperatures fan...

Страница 71: ...associated with this query Parameter Query Response Not applicable NR1 Operation Status Word Bits Definition D0 CALibrating D1 Not Used D2 Not Used D3 Not Used D4 Not Used D5 Not Used D6 Not Used D7 N...

Страница 72: ...0 IntelliFrame Instruction Manual 3 31 Query Response STATus OPERation CONDition STATus OPERation ENABle Examples Related Commands Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SO...

Страница 73: ...e STATus OPERation command on page 3 30 Unused bits always return a value of 0 Reading the Operation Condition register does not clear it There is no command associated with this query Parameter Query...

Страница 74: ...register value at power on is set to H0000 Executing the STATus PRESet command sets the register value to H0000 STATus OPERation ENABle Mask STATus OPERation ENABle Mask Query Response NRf Nondecimal...

Страница 75: ...and associated with this query Parameter Query Response Not applicable NR1 VXI Monitor Event Word Bits Definition D0 BERR Occurred D1 SYSFAIL Occurred D2 ACFAIL Occurred D3 AS Missing D4 Not Used D5 N...

Страница 76: ...em VX1410 IntelliFrame Instruction Manual 3 35 Query Response STATus OPERation VXIMonitor ENABle Examples Related Commands Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www...

Страница 77: ...en no ACFAIL is detected the query returns a value of 0 in all fields At power on the values of all fields are set to 0 There is no command associated with this query Parameter Query Response Not Appl...

Страница 78: ...query returns NONE At power on the value is set to NONE There is no command associated with this query Parameter Query Response Not Applicable A16 A24 A32 NONE Not Applicable None None Query Response...

Страница 79: ...er value at power on is set to H7FFF Executing the STATus PRESet command sets the register value to H7FFF STATus OPERation VXIMonitor ENABle Mask STATus OPERation VXIMonitor ENABle Mask Query Response...

Страница 80: ...When no SYSFAIL is detected the query returns a value of 0 in all fields At power on the values of all fields are set to 0 There is no command associated with this query Parameter Query Response Not...

Страница 81: ...nt Queue is preset to the default value of 499 0 The SCPI Positive Transition Filters are preset to 1 This command does not clear the SCPI event registers or the IEEE 488 2 enable and event registers...

Страница 82: ...o command associated with this query Parameter Query Response Not applicable NR1 Questionable Status Word Bits Definition D0 VOLTage D1 Not Used D2 Not Used D3 Not Used D4 TEMPerature D5 Not Used D6 N...

Страница 83: ...410 IntelliFrame Instruction Manual Query Response STATus QUEStionable CONDition STATus QUEStionable ENABle Examples Related Commands Artisan Technology Group Quality Instrumentation Guaranteed 888 88...

Страница 84: ...The register is cleared by the CLS command There is no command associated with this query Parameter Query Response Not applicable NR1 Blower Status Word Bits Definition D0 Blower 1 D1 Blower 2 D2 Not...

Страница 85: ...Manual Reading the register clears it Query Response STATus QUEStionable BLOWer CONDition STATus QUEStionable BLOWer ENABle Dependencies Examples Related Commands Artisan Technology Group Quality Inst...

Страница 86: ...register can be found under the STA Tus QUEStionable BLOWer command on page 3 43 Reading this register does not clear it There is no command associated with this query Parameter Query Response Not ap...

Страница 87: ...d on page 3 124 at power on the register value will be restored from nonvolatile RAM otherwise the register value at power on is set to H7FFF Executing the STATus PRESet command sets the register valu...

Страница 88: ...e BLOWer SPEed Blower Blower Query Response BLOWer1 BLOWer2 NR1 Not Applicable 224 Illegal Parameter Value You attempted to use an invalid parameter value None Query STAT QUES BLOW SPE BLOW1 Response...

Страница 89: ...under the STATus QUEStionable command on page 3 41 Reading the register does not clear it There is no command associated with this query Parameter Query Response Not Applicable NR1 Not Applicable None...

Страница 90: ...tored from nonvolatile RAM otherwise the register value at power on is set to H0000 Executing the STATus PRESet command sets the register value to H0000 Mask Query Response NRf Nondecimal Numeric 0 N...

Страница 91: ...ery Response NRf 1 N 65535 MIN MAX NR1 1 65535 Not Applicable 222 Data Out of Range You attempted to set the filter cleaning time interval to an illegal value None Command Query Response STATus QUESti...

Страница 92: ...cycle Parameter Query Response Not Applicable NR1 Not Applicable Not Applicable None Query Response STATus QUEStionable FILTer TINTerval STATus QUEStionable FILTer TRESet Syntax Parameters Reset Value...

Страница 93: ...lay when the front panel displays the filter cleaning warning message or the filter cleaning remaining time Option 1M only Parameter Query Response Not Applicable Not Applicable Not Applicable Not App...

Страница 94: ...ts Definition D0 Slot 0 Delta Temperature D1 Slot 1 Delta Temperature D2 Slot 2 Delta Temperature D3 Slot 3 Delta Temperature D4 Slot 4 Delta Temperature D5 Slot 5 Delta Temperature D6 Slot 6 Delta Te...

Страница 95: ...rame Instruction Manual Query Response STATus QUEStionable TEMPerature CONDition STATus QUEStionable TEMPerature ENABle Examples Related Commands Artisan Technology Group Quality Instrumentation Guara...

Страница 96: ...t ambient input temperature OUT0 OUT12 output temperature of each slot DELTa0 DELTa12 output temperature minus input temperature of each slot NR1 Not Applicable 224 Illegal Parameter Value You attempt...

Страница 97: ...Perature command on page 3 53 If the fan speed is already set to the maximum speed and the delta temperature is above the limit for 30 seconds a failure will be reported Reading the register does not...

Страница 98: ...et to H7FFF Executing the STATus PRESet command sets the register value to H7FFF STATus QUEStionable TEMPerature ENABle Mask STATus QUEStionable TEMPerature ENABle Mask Query Response NRf Nondecimal N...

Страница 99: ...0 OUT12 output temperature of each slot DELTa0 DELTa12 output temperature minus input temperature of each slot NR1 Not Applicable 224 Illegal Parameter Value You attempted to use an illegal parameter...

Страница 100: ...e of each slot NR1 Limit NRf If Location is AMBient 1 Limit 55 MIN MAX If Location is DELTan 5 Limit 30 MIN MAX 1 55 5 30 Not Applicable 224 Illegal Parameter Value You attempted to set the location t...

Страница 101: ...a0 DELTa12 AMBient ambient input temperature OUT0 OUT12 output temperature of each slot DELTa0 DELTa12 output temperature minus input temperature of each slot NR1 Not Applicable 224 Illegal Parameter...

Страница 102: ...is cleared by the CLS command There is no command associated with this query Parameter Query Response Not applicable NR1 Voltage Status Word Bits Definition D0 24 V D1 12 V D2 5 V D3 5 V Standby D4 5...

Страница 103: ...anual Reading the register clears it Query Response STATus QUEStionable VOLTage CONDition STATus QUEStionable VOLTage ENABle Dependencies Examples Related Commands Artisan Technology Group Quality Ins...

Страница 104: ...s in this register can be found in STATus QUEStionable VOLTage command on page 3 61 Reading the register does not clear it There is no command associated with this query Parameter Query Response Not A...

Страница 105: ...set to H7FFF Executing the STATus PRESet command sets the register value to H7FFF STATus QUEStionable VOLTage ENABle Mask STATus QUEStionable VOLTage ENABle Mask Query Response NRf Nondecimal Numeric...

Страница 106: ...4 V P12 12 V P5 5 V P5STBY 5 V Standby P5EXT 5 V External N2 2 V N5PT2 5 2 V N12 12 V N24 24 V Not Applicable 224 Illegal Parameter Value You attempted to use an illegal parameter value None Query Res...

Страница 107: ...of the transition filter are always enabled Bit 15 always returns a 0 The factory default setting is 32743 The 5 V Standby and 5 V External are not monitored STATus QUEStionable VOLTage PTR Mask STATu...

Страница 108: ...m the standard SCPI command The SYSTem ER Ror query is an alias There is no command associated with this query Parameter Query Response Not Applicable NR1 string Not Applicable None None Query Respons...

Страница 109: ...CLS command The SYSTem ERRor ALL query is an alias There is no command associated with this query Parameter Query Response Not Applicable NR1 string Not Applicable None None Query Response 221 Setting...

Страница 110: ...s cleared at power on or when you send the CLS command There is no command associated with this query Parameter Query Response Not Applicable NR1 Not Applicable None None Query Response STATus QUEue S...

Страница 111: ...ror Event queue is cleared at power on or when you send the CLS command There is no command associated with this query Parameter Query Response Not Applicable NR1 Not Applicable None None Query Respon...

Страница 112: ...vent queue is cleared at power on or when you send the CLS command There is no command associated with this query Parameter Query Response Not Applicable NR1 Not Applicable None None Query Response ST...

Страница 113: ...n you send the STATus PRESet command the contents of the register is set to 499 0 Event List Query Response NRf numeric range NRf numeric range NRf numeric range NRf numeric range numeric range is NRf...

Страница 114: ...ble to the value of 1999 1000 The SYS Tem ERRor ENABle ALL command is an alias Not Applicable None None Command STATus PRESet STATus QUEue ENABle CLS Syntax Reset Value Errors and Events Dependencies...

Страница 115: ...of zero Definition of the bits can be found in STATus SEVent command on page 3 75 Reading the register does not clear it There is no command associated with this query Parameter Query Response Not Ap...

Страница 116: ...ommand associated with this query Parameter Query Response Not applicable NR1 Event Status Word Bits Definition D0 Blower 1 D1 Blower 2 D2 Filter D3 Slot 0 Delta Temperature D4 Slot 1 Delta Temperatur...

Страница 117: ...ternal D22 2 V D23 5 2 V D24 12 V D25 24 V D26 BERR D27 SYSFAIL D28 ACFAIL D29 AS D30 Not Used D31 Not Used Not Applicable None None Query Response STATus SCONdition Reset Value Errors and Events Depe...

Страница 118: ...1410 IntelliFrame These default settings do not change when you issue the SYSTem FACTory command Table 3 14 Default Factory RS 232 Communication Settings Command Factory Default Setting SYSTem COMMuni...

Страница 119: ...Baud Rate Query Response NRf 300 600 1200 2400 4800 9600 19200 MINimum MAXimum NR1 300 600 1200 2400 4800 9600 19200 300 19200 Not Applicable 222 Data Out of Range You attempted to set the baud rate...

Страница 120: ...y Response NRf 7 8 MINimum MAXimum NR1 7 8 7 8 Not Applicable 222 Data Out of Range You attempted to set the bits per character value to an illegal value None Command Query Response All SYSTem COMMuni...

Страница 121: ...nd ignores the incoming CTS line A value of IBFull or RFR specifies that the two lines are used for hardware flow control Flow Control Query Response OFF ON IBFull RFR OFF ON IBF Not Applicable 224 Il...

Страница 122: ...aracters are echoed back to the RS 232 port Boolean Query Response NRf N 0 0 ON OFF NR1 1 0 1 0 Not Applicable None None Command Query Response All SYSTem COMMunicate SERial commands Syntax Parameters...

Страница 123: ...queue and will not be output to the RS 232 port directly If the mode is ON all asynchronous error messages are placed in the Error Event queue The event registers will not be updated by any event mes...

Страница 124: ...put before the newline character is received For example for a backspace character use H for a line delete character use U If you enable the line buffer mode the output character n 0x0A will be conver...

Страница 125: ...values Avoid using XON pacing when transferring binary data Flow Control Query Response NONE XON NONE XON Not Applicable 224 Illegal Parameter Value You attempted to set the flow control to an illega...

Страница 126: ...ansmit and receive values Type Query Response EVEN NONE ODD EVEN NONE ODD Not Applicable 224 Illegal Parameter Value You attempted to set the parity to an illegal value None Command Query Response All...

Страница 127: ...AW mode There are no associated queries with these commands Table 3 15 lists the preset values for each PRESet command Table 3 15 Effects of PRESet on Serial Port Parameters SERial Port Parameter ALL...

Страница 128: ...th the transmit and receive values Stop Bits Query Response NRf 1 2 MINimum MAXimum NR1 1 2 1 2 Not Applicable 222 Data Out of Range You attempted to set terminator to an illegal value None Command Qu...

Страница 129: ...ere is no command associated with this query Parameter Query Response None year month day Not Applicable None None Query Response SYSTem TIME LMAintenance Syntax Parameters Reset Value Errors and Even...

Страница 130: ...item from the queue Parameter Query Response Not Applicable NR1 string Not Applicable 350 Queue Overflow The Error Event queue overflowed due to the execution errors of other com mands The SYST ERR q...

Страница 131: ...pt of the CLS command There is no associated command for this query SYSTem ERRor ALL Parameter Query Response Not Applicable NR1 string Not Applicable None None Query SYST ERR ALL Response 221 Setting...

Страница 132: ...is cleared at power up upon receipt of the CLS command There is no associated command for this query Parameter Query Response Not Applicable NR1 Not Applicable None None Query Response STATus QUEue S...

Страница 133: ...nd The Error Event queue is cleared at power up upon receipt of the CLS command There is no associated command for this query Parameter Query Response Not Applicable NR1 Not Applicable None None Query...

Страница 134: ...Event queue is cleared at power up upon receipt of the CLS command There is no associated command for this query Parameter Query Response Not Applicable NR1 Not Applicable None None Query Response STA...

Страница 135: ...hen you send the STATus PRESet command the contents of the register is set to 499 0 Event List Query Response NRf numeric range NRf numeric range NRf numeric range NRf numeric range numeric range is N...

Страница 136: ...value of 1999 1000 The STA Tus QUEue ENABle ALL command is an alias Not Applicable None None Command STATus PRESet STATus QUEue ENABle STATus QUEue ENABle ALL Syntax Reset Value Errors and Events Dep...

Страница 137: ...us QUEStionable BLOWer ENABle 32767 STATus QUEStionable ENABle 0 STATus QUEStionable FILTer TINTerval 4000 STATus QUEStionable TEMPerature ENABle 32767 STATus QUEStionable TEMPerature LIMit AMBient 55...

Страница 138: ...m Subsystem VX1410 IntelliFrame Instruction Manual 3 97 None Command None Dependencies Examples Related Commands Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg...

Страница 139: ...query Command Header Query Response string string Not Applicable The query returns a null string if the Command Header is not a valid command header being recognized by the VX1410 IntelliFrame None Q...

Страница 140: ...and from the VXI interface the interface event message is placed in the RS 232 Error Event queue as part of an event message The message in the Error Event queue looks similar to the following message...

Страница 141: ...ge PTR STATus OPERation VXIMonitor ENABle SYSTem SYSReset STATus QUEStionable BLOWer ENABle VXIMonitor TINTerval STATus QUEStionable ENABle VXIMonitor REPeat STATus QUEStionable FILTer TINTerval ESE S...

Страница 142: ...PACE STATus QUEStionable ENABle SYSTem COMMunicate SERial PARity STATus QUEStionable FILTer TINTerval SYSTem COMMunicate SERial SBITs STATus QUEStionable TEMPerature ENABle SYSTem SYSReset STATus QUE...

Страница 143: ...on or off Boolean Query Response NRf N 0 0 ON OFF NR1 1 0 1 0 Not Applicable 224 Illegal Parameter Value You attempted to set the parameter to an illegal value None Command Query Response None Syntax...

Страница 144: ...wer has been cycled since the mainframe was manufactured Parameter Query Response Not Applicable NR1 Not Applicable None None Query Response None Syntax Parameters Reset Value Errors and Events Depend...

Страница 145: ...from the mainframe power supply or from an external power supply Parameter Query Response Not Applicable MAIN EXTERNAL Not Applicable None None Query Response None Syntax Parameters Reset Value Errors...

Страница 146: ...ET is enabled and the mainframe is powered down commu nication through the RS 232 port is not possible even if an external 5 V is supplied to the enhanced monitor card The display will not show the cu...

Страница 147: ...ce the last factory maintenance There is no command associated with this query Parameter Query Response None hour minute second Not Applicable None None Query Response SYSTem TIME ON Syntax Parameters...

Страница 148: ...wer on There is no command associated with this query Parameter Query Response Not Applicable hour minute second Not Applicable None None Query Response SYSTem TIME LMAintenance Syntax Parameters Rese...

Страница 149: ...Frame complies There is no command associated with this query Parameter Query Response Not Applicable NR2 1995 0 Not Applicable None None Query Response IDN Syntax Parameters Reset Value Errors and Ev...

Страница 150: ...nd DAC Control 1 Yes Yes Yes 1100 Front Panel Display Control Test 1 25 No No Yes 1110 Front Panel Display Control Test 2 36 No No Yes 1120 Front Panel Display Control Test 3 6 No No Yes A description...

Страница 151: ...ixels H Front Panel Display Control Test 2 This test writes the following ASCII characters to the display 0x20 to 0x7E These characters are followed by and c H Front Panel Display Control Test 3 This...

Страница 152: ...res The command form executes the same tests but returns no results code Each internal self test has a unique numeric identifier in the range of 1000 to 1999 see Table 3 19 on page 3 109 TST and TEST...

Страница 153: ...failures Each internal self test has a unique numeric identifier in the range of 1000 to 1999 see Table 3 19 on page 3 109 Parameter Query Response NRf Non decimal number 1000 N 1999 NR1 0 1000 N 1999...

Страница 154: ...no failures There is no associated command for this query Each internal self test has a unique numeric identifier in the range of 1000 to 1999 see Table 3 19 on page 3 109 Parameter Query Response Not...

Страница 155: ...ted A value of zero is returned if there were no failures There is no associated command for this query Parameter Query Response Not Applicable string error number NR1 0 1000 N 1999 verbose message er...

Страница 156: ...ble the repeat mode and use the query to return the status of the repeat mode The VXIMonitor subsystem controls the parameters shown in Figure 3 10 Figure 3 10 VXIMonitor Subsystem Hierarchy The VXIMo...

Страница 157: ...onitors the VXI bus for the specified time interval The factory default setting is OFF 0 Boolean Query Response NR1 N 0 0 ON OFF NR1 1 0 1 0 Not Applicable 221 Settings Conflict You attempted to set t...

Страница 158: ...lean Query Response NR1 N 0 0 ON OFF NR1 1 0 1 0 0 213 Init Ignored You sent INITiate while the VXI Monitor was already monitoring the VXI bus None Command Query Response VXIMonitor TINTerval VXIMonit...

Страница 159: ...setting is 10 seconds time Query Response NRf 1 N 2147483647 MIN MAX NR1 1 2147483647 Not Applicable 222 Data Out of Range You attempted to set the interval to an illegal value 221 Settings Conflict Y...

Страница 160: ...and IEEE 488 2 event registers and the Error Event queue There is no query associated with this command SCPI and IEEE 488 2 enable registers are not cleared by this command None Not Applicable None N...

Страница 161: ...HFF NR1 Standard Status Byte Definition Query Response D0 D1 D2 D3 D4 D5 D6 D7 Operation Complete Not Used Query Error Device Dependent Error Execution Error Command Error Not Used Power On Not Appli...

Страница 162: ...er Query Response Not Applicable NR1 Standard Status Byte Definition Query Response D0 D1 D2 D3 D4 D5 D6 D7 Operation Complete Not Used Query Error Device Dependent Error Execution Error Command Error...

Страница 163: ...age includes both the SCPI and firmware version numbers Parameter Query Response Not Applicable Syntax Defined by IEEE 488 2 Manufacturer Model Number Serial Number Firmware Version or Equivalent Not...

Страница 164: ...tatus register after the previous command has completed execution The query causes an ASCII one to be placed in the output queue after the previous command has completed execution None Not Applicable...

Страница 165: ...Status Enable Register and the SCPI Enable Register settings will be restores to their previous settings Otherwise the settings will be cleared The factory default setting is 0 Value Query Response NR...

Страница 166: ...command This command has no effect other than stopping the VXI Monitor The VXI Monitor stops execution while other monitors remain active None Not Applicable None None Command STATus PRESet CLS Synta...

Страница 167: ...is is not required by IEEE 488 2 The factory default setting is 0 Mask Query Response NRf Non Decimal Numeric 0 N HFF NR1 Standard Status Byte Definition Query Response D0 D1 D2 D3 D4 D5 D6 D7 Not Use...

Страница 168: ...R1 Standard Status Byte Definition Definition D0 D1 D2 D3 D4 D5 D6 D7 Not Used Not Used Error Event Queue Questionable Event Register MAV ESB Standard Event Status Reg RQS MSS Operation Event Register...

Страница 169: ...al self test has a unique identifier in the range of 1000 to 1999 TST and TEST are identical queries There is no command associated with this query You can obtain additional test results with the TEST...

Страница 170: ...ery associated with this command WAI prevents the instrument from executing further commands until the previous command has completed execution None Not Applicable None None Command OPC Syntax Paramet...

Страница 171: ...nteger with the most significant bit always zero positive logic Unused or don t care bits in masks and registers always return zeros when queried The Operation Status register and the Questionable Sta...

Страница 172: ...UEStionable EVENt STATus QUEStionable CONDition STATus QUEStionable ENABle Error Event Queue Not Empty Summary of VOLTage FILTer Summary of TEMPerature Summary of BLOWer Summary of VXI Monitor Power O...

Страница 173: ...example one of the power supplies is not working properly you could program the mainframe to warn you of possible problems There are three status registers that provide summary information to the Que...

Страница 174: ...esponding bit in the Temperature Status register If a problem occurs the corresponding bit will be set in the register 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 TEMPerature Status Registers Not Used DELTA...

Страница 175: ...nd event registers and two separate Error Event queues for the two interfaces You can send the commands listed below from the interface to set up only those status and event registers of that interfac...

Страница 176: ...ntelliFrame execution control block Execution errors can occur when a program data element following a header was evaluated as outside of the legal input range or when the program data element is inco...

Страница 177: ...data was invalid for the header 144 Character data too long The character element was more than 12 characters long 160 Block data error 161 Invalid block data An invalid block data element was receive...

Страница 178: ...y INTERRUPTED The query was interrupted before a response was completely sent 420 Query UNTERMINATED An incomplete program message was received 430 Query DEADLOCKED The input and output buffers are fu...

Страница 179: ...he bit 0 in the Questionable Enable register If any problems occur with any of the power supplies bit 0 of the register will be set Example 2 Monitor the Delta Temperature This example shows how to mo...

Страница 180: ...tion The following example shows how to return a summary of the status information using the STAT SEVent command Send the STAT SEVent query and note the returned information For example assume that th...

Страница 181: ...MS continuous range Source Frequency 45 Hz to 440 Hz continuous range 115 V operation 45 Hz to 66 Hz 230 V operation Power Consumption 1350 W line power maximum Fuse Rating 90 V 132 V Operation 0 25 i...

Страница 182: ...A maximum if user supplied Maximum Current Capacity IMD Voltage IMD dynamic current Specification only valid for Main ame se ial m e s 020100 a 24 V 6 5 A frame serial numbers B020100 and higher 12 V...

Страница 183: ...ough VXIbus or RS 232 with Option 1M Voltage Readout Accuracy typical 3 1 If the VX1410 IntelliFrame is configured with Option 1D Delete Power Supply all product specifications in this table do not ap...

Страница 184: ...ar of mainframe Mainframe Exhaust Top sides of mainframe also on top rear of mainframe Cooling Clearance Two inches on both sides and rear of mainframe Pressurization Positive modules pressurized Filt...

Страница 185: ...through module l s 1 25 1 00 0 75 0 50 0 25 0 00 0 2 4 6 8 10 12 14 16 Pressure across module mm H2O Note Proposed VXI 8 Cooling Methods Figure 4 1 Mainframe Cooling Curve Artisan Technology Group Qu...

Страница 186: ...gured with Option 1D Delete Power Supply all product specifications in this table do not apply To be VXI compliant the power supply must adhere to the specifications of the VXI standard 2 If the VX141...

Страница 187: ...ft 305 m above 5000 ft 1524 m Nonoperating 40 000 ft 12 190 m Dynamics Random Vibration Three axis 30 minutes total 10 minutes per axis Operating 0 27 gRMS total from 5 Hz to 500 Hz Nonoperating 2 28...

Страница 188: ...15 kV with no loss of stored data or damage that prevents recovery of normal operation by the user Immunity Fast Transients Com mon Mode No loss of stored data change to control settings degradation...

Страница 189: ...anical Characteristic Description Overall Dimensions Standard Mainframe Height with feet 14 4 in 36 58 cm Width 16 7 in 42 55 cm Depth 24 in 60 96 cm Mainframe with Rackmount Height 14 0 in 35 56 cm W...

Страница 190: ...inframe 100 lbs 45 4 kg maximum with no modules installed Cable Tray Option 1U 8 lbs 3 6 kg Cable Tray Option 2U 8 lbs 3 6 kg Module Size 13 C size VXI slots The mainframe also accepts A size or B siz...

Страница 191: ...0 0 cm 15 75 in 44 5 cm 17 50 in 48 0 cm 18 9 in 67 06 cm 26 4 in Min 79 76 cm 31 4 in Max Figure 4 3 Front View and Side View of VX1410 IntelliFrame with Option 1U and Rackmount Option Artisan Techno...

Страница 192: ...Option 2U and Rackmount Option Table 4 10 VXI and Plug Play Characteristic Description VXI Fully compatible with VXIbus System Specifications Rev 1 4 Plug Play Mainframe VXI Plug Play compliant with t...

Страница 193: ...TDS500A series oscilloscope with a P6101B probe DC Voltage Regulation The DC voltage regulation can be verified by following the steps listed under Functional Check beginning on page 1 25 If your main...

Страница 194: ...a ground lead less than one inch to the 1X oscilloscope probe P6101B 3 Connect the probe to Pin 1 of the 25 pin connector and connect the ground lead to one of the ground pins listed in Table 5 1 4 C...

Страница 195: ...d injury do not perform any servicing other than that stated in the operating instructions unless you are qualified to do so Refer to all Safety Summaries before performing any service WARNING Artisan...

Страница 196: ...the board by looking through the exhaust holes on the side of the mainframe If the Temperature Sense board is located toward the rear of the mainframe you must remove the mainframe cover and move the...

Страница 197: ...a Bits 8 Stop Bits 1 Parity None Flow Control Xon Xoff 10 Make sure that the mainframe has had at least five minutes to warm up before continuing this procedure 11 Send the following commands to enabl...

Страница 198: ...truction Manual 6 3 17 This completes the calibration procedure The rest of this section describes the CALibration and SYSTem PASSword commands in detail Artisan Technology Group Quality Instrumentati...

Страница 199: ...A return value of 0 means that the calibration has passed and a return value between 10 and 22 indicates a calibration failure in slots 0 through 12 Parameter Query Response Not Applicable NR1 0 10 22...

Страница 200: ...mmand the VX1410 IntelliFrame will generate an error Password Query Response string VX1410 Not Applicable Not Applicable 221 Settings Conflict You attempted to send an invalid password None Command SY...

Страница 201: ...d otherwise the VX1410 IntelliFrame will generate an error Password Query Response string VX1410 Not Applicable Not Applicable 221 Settings Conflict You attempted to send an invalid password None Comm...

Страница 202: ...is abled Enabled commands return a value of 1 disabled return a value of 0 Parameter Query Response Not Applicable NR1 1 0 Not Applicable None None Query Response SYSTem PASSword CDISable Syntax Param...

Страница 203: ...t start with an alphabet ical character followed by an optional alphabetical character and by a maximum of six digits string Not Applicable 221 Settings Conflict You attempted to execute a command tha...

Страница 204: ...he location of controls indicators and connectors used with the mainframe CAUTION Many components within the mainframe are susceptible to static discharge damage Service the mainframe only in a static...

Страница 205: ...ng of the mainframe and cleaning the fan filter Periodic cleaning reduces instrument breakdown and increases reliability Clean the mainframe as needed based on the operating environment Refer to your...

Страница 206: ...ame The filter slides out of the rear of the mainframe as shown in Figure 7 1 Fan Filter Figure 7 1 Removing the Fan Filter Clean the fan filter by vacuuming or blowing air through the filter Rinse th...

Страница 207: ...to Figure 7 3 and Figure 7 4 while perform ing the following steps 1 From the back of the mainframe remove the four 8 32 screws using a 3 32 inch Allen wrench refer to Figure 7 2 for the screw locati...

Страница 208: ...atch the labels on the fan cables to their respective connector on the rear of the mainframe Chassis Ground 8 32 Screw not installed 8 32 Hex Screw Safety Ground 8 32 Screw 8 32 Hex Screw 8 32 Hex Scr...

Страница 209: ...r Supply Handle 8 32 Screws 5 Fan Cables Ledge Fan Assembly Figure 7 3 Removing the Fan Assembly Cable Tie Fan Assembly Fans 6 32 Screws 6 Figure 7 4 Removing the Fans Artisan Technology Group Quality...

Страница 210: ...onitor board Option 1M or 2M remember to calibrate the mainframe Refer to Chapter 5 Adjustment Procedures beginning on page 6 1 for information on calibrating the mainframe To remove the power supply...

Страница 211: ...is normally not necessary because the Power Supply and the Power Supply Interface board can be replaced as a single unit 3 Refer to Figure 7 6 and remove the Power Supply Interface board from the powe...

Страница 212: ...er to pry up the tab of the card guide at the front of the mainframe being careful not to damage the card guide or the mainframe 2 Gently pull the card guide forward until it pops out of place 3 Remov...

Страница 213: ...ward until it pops out of place Slide the card guides toward the rear of the mainframe and allow the front of the card guide to snap into place Figure 7 7 Removing the Top and Bottom Card Guides Artis...

Страница 214: ...2 Remove the 12 screws holding the top cover to the mainframe using a 3 32 inch Allen wrench 3 Slide the top cover of the mainframe back far enough to expose the hole just inside the top and bottom e...

Страница 215: ...the holes at the top of the mainframe being careful not to damage any components on the circuit board NOTE If you replace the Temperature Sense board and you have Option 1M or 2M remember to calibrate...

Страница 216: ...from the rear of the Display assembly 3 Remove the 6 32 screws holding the Display assembly in place 4 Gently remove the Display assembly from the mainframe When you reconnect the ribbon cable to the...

Страница 217: ...lly or J23 Remember to reconnect the cable to the correct connector 3 Disconnect the power switch cable at J22 or from J23 at the top of the backplane 4 From the rear of the mainframe remove the five...

Страница 218: ...ntelliFrame Instruction Manual 7 15 BackPlane Connector 6 32 Screws 17 Power Switch Connector Figure 7 11 Removing the Backplane Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOUR...

Страница 219: ...eration the line fuse and a 5 V fuse on the backplane The line fuse can be one of three different kinds depending on the operating voltage of the mainframe Table 7 1 summarizes the different fuses in...

Страница 220: ...oling Cooling problems can be due to one or more of the following problems H Verify that the air filter is clean If not remove the air filter and clean it H One or both fans may not be operating prope...

Страница 221: ...nections to module slots 0 through 12 in the mainframe It also has connections to all other circuit boards and modules in the mainframe The front panel On Standby DC switch connects to one of two conn...

Страница 222: ...n Standby Switch External Power Logical Addr RS232 Passive Monitor Fan Speed Switch P1 P2 0J4 0P3 0J1 0J2 J23 J22 3 3 J20 J21 J24 0P2 0P1 1P1 2P1 3P1 4P1 5P1 6P1 7P1 8P1 9P1 10P1 11P1 12P1 1P2 2P2 3P2...

Страница 223: ...board includes a nine pin RS 232 port to connect to an external host Address switches near the bottom rear of the board allow you to select the VXIbus address A green READY indicator and amber ACCESS...

Страница 224: ...mber if applicable If you order a part that has been replaced with a different or improved part your local Tektronix field office or representative will contact you concerning any change in part numbe...

Страница 225: ...laceable Parts Name Description Tektronix Part Number Power Supply Module Power supply and power supply interface board 020 2116 XX Air Filter Air filter Woven Polypropylene black 0 125 thick 2 layer...

Страница 226: ...ive Column four indicates the serial number at which the part was discontinued No entries indicates the part is good for all serial numbers 5 Qty This indicates the quantity of parts used 6 Name Descr...

Страница 227: ...RD DRIVE BEAVERTON OR 97006 22589 ELECTRO SPACE FABRICATORS INC 300 W HIGH ST PO BOX 67 TOPTON PA 19562 1420 3M099 PORTLAND SCREW CO 6520 N BASIN ST PORTLAND OR 97217 3920 4T165 NEC ELECTRONICS INC 47...

Страница 228: ...32 X 0 375 PNH STL CDPL T 20 0KB01 ORDER BY DESC STANDARD ACCESSORIES 040 1480 XX 1 FIELD KIT ENHANCED MONITOR UPGRADE OPTION 1M 80009 040 1480 XX 040 1581 XX 1 FIELD KIT ENHANCED MONITOR UPGRADE OPT...

Страница 229: ...able Parts VX1410 IntelliFrame Instruction Manual 8 7 1 2 3 4 5 6 7 4 7 6 5 Figure 8 1 Cabinet and Chassis Assembly Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisa...

Страница 230: ...X 0 500 PNH STL CDPL T 15 0KB01 ORDER BY DESC 7 118 9142 00 1 BACKPLANE VXI 13 SLOT C SRE W ELECTRONIC AUTOMATIC BUS GRANTSENSING 61081 V1213J1J2X 7059 8 213 0882 00 12 SCREW TPG TR 6 233 X 0 437 PNH...

Страница 231: ...VX1410 IntelliFrame Instruction Manual 8 9 1 2 3 4 5 6 7 8 9 10 11 12 13 14 9 8 3 Figure 8 2 Circuit Boards and Chassis Parts Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE...

Страница 232: ...DPL T 20 0KB01 ORDER BY DESC 11 671 3216 00 B010100 B019999 1 CIRCUIT BD ASSY POWER SUPPLY INTERFACE 80009 671321600 11 671 3216 01 B020100 1 CIRCUIT BD ASSY POWER SUPPLY INTERFACE 80009 671321601 12...

Страница 233: ...liFrame Instruction Manual 8 11 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 16 10 10 17 18 14 19 20 21 Figure 8 3 Power Supply and Fan Assembly Artisan Technology Group Quality Instrumentation Guaranteed 8...

Страница 234: ...T CONTAINS ITEMS 1 THROUGH 5 80009 020211700 1 950 0991 00 2 HANDLE ALUMINUM BLK 80009 950099100 2 212 0157 00 14 SCREW MACHINE 8 32 X 0 5 FLH 100 DEG STL CDPL TORX 0KB01 ORDER BY DESC 3 407 4363 00 2...

Страница 235: ...laceable Parts VX1410 IntelliFrame Instruction Manual 8 13 1 2 3 4 5 2 3 2 2 5 Figure 8 4 1R Rackmount Assembly Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg...

Страница 236: ...X 0 5 FLH 100 DEG STL CDPL TORX 0KB01 ORDER BY DESC 3 214 4678 00 2 HINGE LIFT OFF BLACK 80009 214467800 4 407 4363 00 2 BRACKET RACKMOUNT 2R VX1410 80009 407436300 5 211 0718 00 8 SCREW MACHINE 6 32...

Страница 237: ...VX1410 IntelliFrame Instruction Manual 8 15 1 2 3 4 5 6 7 8 9 6 9 4 1 5 2 2 2 8 3 Figure 8 5 2R Rackmount and Door Assembly Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE w...

Страница 238: ...E TRAY 6 0 020 2119 00 1 COMPONENT KIT CABLE TRAY 1U KIT KIT INCLUDES ITEMS 1 THROUGH 3 80009 020211900 1 212 0158 00 2 SCREW MACHINE 8 32 X 0 375 PNH STL CDPL T 20 0KB01 ORDER BY DESC 2 333 4158 00 1...

Страница 239: ...Replaceable Parts VX1410 IntelliFrame Instruction Manual 8 17 1 2 3 Figure 8 6 Option 1U Cable Tray Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Страница 240: ...T KIT CAB E TRAY 2U KIT VX1410 KIT INCLUDES THE ITEMS 1 THROUGH 4 80009 020212000 1 212 0158 00 4 SCREW MACHINE 8 32 X 0 375 PNH STL CDPL T 20 0KB01 ORDER BY DESC 2 333 4162 00 1 PANEL CONN 2U 0 09 AL...

Страница 241: ...Replaceable Parts VX1410 IntelliFrame Instruction Manual 8 19 1 2 3 4 Figure 8 7 Option 2U Cable Tray Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Страница 242: ...5301 7 211 0501 00 2 SCREW MACHINE 6 32 X 0 125 PNH STL CD PL POZ 73893 ORDER BY DESC 346 0128 00 1 STRAP TIEDOWN E 0 098 W X 4 0 L ZYTEL TK1499 HW 047 8 407 4332 00 1 BRACKET 0 062 ALUM SHEET METAL M...

Страница 243: ...VX1410 IntelliFrame Instruction Manual 8 21 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 Figure 8 8 Option 1M Enhanced Monitor Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE w...

Страница 244: ...EW MACHINE 6 32 X 0 125 PNH STL CD PL POZ 73893 ORDER BY DESC 346 0128 00 1 STRAP TIEDOWN E 0 098 W X 4 0 L ZYTEL TK1499 HW 047 3 407 4332 00 1 BRACKET 0 062 ALUM SHEET METAL MOUNTS ENHANCED MONITOR E...

Страница 245: ...aceable Parts VX1410 IntelliFrame Instruction Manual 8 23 1 2 3 4 5 6 7 8 Figure 8 9 Option 2M Enhanced Monitor Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg...

Страница 246: ...Description Mfr Code Mfr PartNumber OPTION 1D DELETE POWER SUPPLY 441 4402 00 1 BRACKET BRACKET PSIF DUMMY P S VX1410 NOT ILLUSTRATED 80009 441440200 211 0846 00 4 SCREW SCREW SHLDR 6 32 X 0 156 HEX S...

Страница 247: ...dynamic current IMD for each module the dynamic current is only valid for mainframes with serial number B020100 or higher Add the individual currents to determine the total current needed for each po...

Страница 248: ...1 2 3 4 5 6 7 8 9 10 11 12 VX1410 Current Limit Current Limits Power Calculation Total Power 872W 80A 12A 13A 3A 13A 3A 12A 6 5A 12A 6 5A 60A 9A 30A 5A Individual Current Sums Less Than VX1410 Total...

Страница 249: ...pendix Remote Power Switch Configuration You can control the power on function of the mainframe by the front panel On Standby switch and from a remote momentary switch through the connec tions of the...

Страница 250: ...place the mainframe cover 6 Connect the momentary switch to pins 5 and 18 return side of the 25 pin rear panel connector Power Switch Connector Backplane J22 J23 Figure B 1 Power Switch Connectors Art...

Страница 251: ...er location on the Standard Monitor board The jumper is shipped in the 8 10 position You can move the jumper to the 13 15 position if there is a high temperature rise expected in the modules or when t...

Страница 252: ...ltage from the 5 V external supply Monitor Backup Power Supply connector then the external supply will provide the backplane with the 5 V external voltage 1 A maximum If you source the 5 V Standby vol...

Страница 253: ...necessary design the power supply cables so that they can be routed through the AC power connector on the Fan Assembly CAUTION To prevent damage to the mainframe do not operate the mainframe with the...

Страница 254: ...a connector to connect to the P1 and P2 Power Supply Interface connectors Figure B 5 shows an example of the connector You can assemble a mating connector using parts from the AMPOWER Wave Crimp Syst...

Страница 255: ...2 V Sense RTN 2 V Sense 5 V 24 V RTN 24 V No Connection 2 V Sense RTN 5 V Sense 5 V Sense RTN 12 V 2 V RTN 12 V ACFAIL RTN No Connection On Standby RTN SYSRESET RTN 5 2 V RTN No Connection Figure B 5...

Страница 256: ...commands supported by the VX1410 IntelliFrame The columns at right show whether a command is defined in the SCPI 1995 0 Standard or not Table C 1 SCPI Conformance Information Command Defined in SCPI 1...

Страница 257: ...ction Manual Table C 1 SCPI Conformance Information Cont Command Not Defined In SCPI 1995 0 Defined in SCPI 1995 0 n n n n n n n n n n n n n n n n n n n n n n n n n n n n n n n n Artisan Technology Gr...

Страница 258: ...n ENABle n ALL n FACTory n HELP SYNTax n IEMessage n NVRecall n NVSave n PASSword CDISable n CENable n STATe n POWer n CYCLe n SOURce n SYSReset n TIME LMAintenance n ON n VERSion n TEST ALL n NUMber...

Страница 259: ...X1410 IntelliFrame Instruction Manual Table C 1 SCPI Conformance Information Cont Command Not Defined In SCPI 1995 0 Defined in SCPI 1995 0 n n n n n n n Artisan Technology Group Quality Instrumentati...

Страница 260: ...as a VMEbus single height short module Asynchronous Communication Communications that occur outside the normal command response cycle Such communications have higher priority than synchronous communi...

Страница 261: ...timing skew of 2 ns and a maximum total delay of 8 ns Commander In the VXIbus interface a device that controls another device a servant A commander may be a servant of another commander Command A dire...

Страница 262: ...is not defined in the VXIbus specification D Size Card A VXIbus instrument module that is 340 0 366 7 mm 30 48 mm 13 4 14 4 in 1 2 in DTB See Data Transfer Bus DTB Arbiter A functional module that ac...

Страница 263: ...ce Also one device may require more than one instrument module IntelliGuides The card guides used in the VX1410 IntelliFrame The card guides on the bottom of the mainframe shut off airflow when no mod...

Страница 264: ...a well defined terminator and message body Message Based Device A VXIbus device that supports VXI configuration and communication registers Such devices support the word serial protocol and possibly...

Страница 265: ...system hierarchy allocating shared system resources performing system self test diagnostics and initializing system commanders Self Calibration A routine that verifies the basic calibration of the in...

Страница 266: ...e lines are well matched for timing skew STST STart STop protocol used to synchronize modules SYNC100 A Slot 0 signal that is used to synchronize multiple devices with respect to a given rising edge o...

Страница 267: ...at performs slot 0 functions and a given complement of instrument modules The subsystem may also include a Resource Manager Word Serial Protocol A VXIbus word oriented bi directional serial protocol f...

Страница 268: ...4 5 D device specific error 3 136 diagnostics 7 16 disassembly procedures 7 4 7 15 tools required 7 4 display assembly removing 7 13 Display board 7 20 DISPlay Subsystem 3 27 3 28 DISPlay WINDow TEXT...

Страница 269: ...monitor display assembly removing 7 13 N nut rails 7 11 removing 7 11 O On Standby switch 1 12 2 2 7 18 B 1 OPC 3 123 Operation Status register 3 131 Operation Status Word 3 30 P packaging 7 17 parity...

Страница 270: ...EStionable FILTer TRESet 3 52 STATus QUEStionable TEMPerature 3 53 3 54 STATus QUEStionable TEMPerature CMAXi mum 3 55 STATus QUEStionable TEMPerature CONDition 3 56 STATus QUEStionable TEMPerature EN...

Страница 271: ...NUMBer 3 112 TEST RESults 3 113 TEST RESults VERBose 3 114 theory of operation 7 18 7 20 tree hierarchy depicted 3 22 troubleshooting procedures 7 16 7 17 equipment required 7 16 TST 3 128 U upgrade k...

Страница 272: ...uipment Have surplus equipment taking up shelf space We ll give it a new home Learn more Visit us at artisantg com for more info on price quotes drivers technical specifications manuals and documentat...