Commands Listed in Alphabetical Order
MEASUREMENT:MEAS2:SOURCE1:GATE2:PCTPOS?
might return
MEASUREMENT:MEAS2:SOURCE1:GATE2:PCTPOS 8.00000001
,
indicating that Gate 2 is set at 80% on the Source 1 waveform used for
Measurement 2.
MEASUrement:MEAS<x>:SOUrce<x>:GATE<x>:POS
This command sets or queries the position (in waveform units) of the source gates
for the measurement. The measurement, source, and gate are de
fi
ned as follows:
MEAS<x> where x can be 1 through 8 (a maximum of 8 measurement slots)
SOURCE<x> where x can be 1 or 2 (a maximum of 2 source waveforms)
GATE<x> where x can be 1 or 2 (one pair of gates per source waveform)
Gates are generic, nondirectional markers that segment the area of waveform
data over which measurements are performed. There is no distinction in gates of
stop or start, so they can cross.
Although you can use this command to set the position of a gate in waveform
units, a user cannot directly set the position of a gate in the Measurement Setup
dialog box using waveform units. Instead, the user selects the Region dialog,
enters the position as a percentage of the waveform, and then reads the position
in waveform units in the readouts adjacent to the percentage controls for G1
and G2. This is equivalent to setting the position with the MEASUrement
MEAS<x>:SOURCE<x>:GATE<x>:PCTPOS command.
NOTE.
The SOURCE<x> argument speci
fi
es the source, which can be source 1
or source 2. To set the actual waveform (channel, reference, or math) that is
assigned to source, you must use the MEASUrement:MEAS<x>:SOURCE<x>
WFM command. Two sources are provided because some measurements (gain,
phase) require two sources.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:SOUrce<x>:GATE<x>:POS <NR3>
MEASUrement:MEAS<x>:SOUrce<x>:GATE<x>:POS?
Related Commands
MEASUrement:MEAS<x>:SOUrce<x>:GATE<x>:PCTPos
MEASUrement:MEAS<x>:GATing:STATE
Arguments
NR3
is the position of the speci
fi
ed gate in waveform units. The position is
constrained to be
≥
TOFP (time of
fi
rst point) and
≤
TOLP (time of last point).
2-250
DSA8300 Programmer Manual
Содержание DSA8300
Страница 1: ...xx DSA8300 Digital Serial Analyzer ZZZ Programmer Manual P077057002 077 0570 02 ...
Страница 2: ......
Страница 3: ...DSA8300 Digital Serial Analyzer ZZZ Programmer Manual xx Revision B www tektronix com 077 0570 02 ...
Страница 8: ...Preface iv DSA8300 Programmer Manual ...
Страница 9: ...Getting Started ...
Страница 10: ......
Страница 15: ...Syntax and Commands ...
Страница 16: ......
Страница 404: ...Commands Listed in Alphabetical Order 2 388 DSA8300 Programmer Manual ...
Страница 405: ...Status and Events ...
Страница 406: ......
Страница 426: ...Status and Events 3 20 DSA8300 Programmer Manual ...
Страница 427: ...Examples ...
Страница 428: ......
Страница 431: ...Appendices ...
Страница 432: ......
Страница 433: ...Appendix A Character Set DSA8300 Programmer Manual A 1 ...
Страница 434: ...Appendix A Character Set A 2 DSA8300 Programmer Manual ...
Страница 446: ...Appendix C Factory Default Setup Values C 8 DSA8300 Programmer Manual ...