Commands Listed in Alphabetical Order
NOTE.
The SOURCE<x> argument speci
fi
es the source, which can be source 1
or source 2. To set the actual waveform (channel, reference, or math) that is
assigned to source, you must use the MEASUrement:MEAS<x>:SOURCE<x>
WFM command. Two sources are provided because some measurements (gain,
phase) require two sources.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:SOUrce<x>:HILow:TRACk:HIGH:ENABle { ON
| OFF | NR1 }
MEASUrement:MEAS<x>:SOUrce<x>:HILow:TRACk:HIGH:ENABle?
Related Commands
MEASUrement:MEAS<x>:SOUrce<x>:HILow:METHod
MEASUrement:MEAS<x>:SOUrce<x>:HILow:TRACk:LOW:ENABle
Arguments
ON
enables tracking of the high value for the speci
fi
ed measurement on the
speci
fi
ed source waveform.
OFF
disables tracking of the high value for the speci
fi
ed measurement on the
speci
fi
ed source waveform.
NR1
set to 0 disables tracking of the high value for the speci
fi
ed measurement
on the speci
fi
ed source waveform; any other value enables tracking of the
high value for the speci
fi
ed measurement on the speci
fi
ed source waveform.
Examples
MEASUREMENT:MEAS2:SOURCE1:HILOW:TRACK:HIGH:ENABLE ON
enables
tracking of the high value used to take Measurement 2 on the Source 1 waveform.
MEASUREMENT:MEAS3:SOURCE2:HILOW:TRACK:HIGH:ENABLE?
might return
MEASUREMENT:MEAS3:SOURCE2:HILOW:TRACK:HIGH:ENABLE 0
, indicating
that the tracking of the high value for Measurement 3 on the Source 2 waveform
is disabled.
NOTE.
A query always returns a 0 or 1.
MEASUrement:MEAS<x>:SOUrce<x>:HILow:TRACk:HIGH:VALue
This command sets or queries the high value used to calculate a speci
fi
ed
measurement on a speci
fi
ed source waveform. The measurement slot is speci
fi
ed
by x, which can be 1 through 8; the source is speci
fi
ed by x, which can be 1 or
2. Sending this command is equivalent to selecting a measurement slot (such as
Meas 3), selecting a source (such as Source1), turning tracking High off, and then
entering an absolute value in the High control box in the in the Hi/Low pane of the
2-254
DSA8300 Programmer Manual
Содержание DSA8300
Страница 1: ...xx DSA8300 Digital Serial Analyzer ZZZ Programmer Manual P077057002 077 0570 02 ...
Страница 2: ......
Страница 3: ...DSA8300 Digital Serial Analyzer ZZZ Programmer Manual xx Revision B www tektronix com 077 0570 02 ...
Страница 8: ...Preface iv DSA8300 Programmer Manual ...
Страница 9: ...Getting Started ...
Страница 10: ......
Страница 15: ...Syntax and Commands ...
Страница 16: ......
Страница 404: ...Commands Listed in Alphabetical Order 2 388 DSA8300 Programmer Manual ...
Страница 405: ...Status and Events ...
Страница 406: ......
Страница 426: ...Status and Events 3 20 DSA8300 Programmer Manual ...
Страница 427: ...Examples ...
Страница 428: ......
Страница 431: ...Appendices ...
Страница 432: ......
Страница 433: ...Appendix A Character Set DSA8300 Programmer Manual A 1 ...
Страница 434: ...Appendix A Character Set A 2 DSA8300 Programmer Manual ...
Страница 446: ...Appendix C Factory Default Setup Values C 8 DSA8300 Programmer Manual ...