Commands Listed in Alphabetical Order
HISTHits
sets the instrument to stop acquisition after it acquires a speci
fi
ed
number of valid hits (that is, non-null samples) in the histogram region. This
is a greater than or equal to condition because only complete records are
processed. Therefore, the number of actual hits will not necessarily match
your requested number of hits; the acquisition will stop when the number
of hits is greater than or equal to the number of speci
fi
ed hits. Use the
ACQuire:STOPAfter:COUNt command to set the speci
fi
ed number of hits.
MASKWaveform
sets the instrument to stop acquisition after a speci
fi
ed
number of the Mask source waveforms are acquired. Use the
ACQuire:STOPAfter:COUNt command to set the speci
fi
ed number of
waveforms.
MASKSamples
sets the instrument to stop acquisition after a greater than or
equal to number of valid (that is, non-null) waveform samples are acquired of
the Mask source waveform. Use the ACQuire:STOPAfter:COUNt command
to set the speci
fi
ed number of waveform samples.
MASK<x>Hits
sets the instrument to stop acquisition after a speci
fi
ed number
of mask hits occur in mask number <x>, which can be 1 through 8. For the
same reasons as HISTHits argument, this is also a greater than or equal to
condition. Use the ACQuire:STOPAfter:COUNt command to set the speci
fi
ed
number of hits.
MASKTOTalhit
sets the instrument to stop acquisition after a speci
fi
ed
number of total mask hits occurs in all masks combined. For the same reasons
as the HISTHits argument, this is also a greater than or equal to condition.
Use the ACQuire:STOPAfter:COUNt command to set the speci
fi
ed total
number of hits.
FRAMecycle
sets the instrument to stop acquisition after it has acquired
the number of bits speci
fi
ed by the HORizontal:FRAMescan:SCANBits
command.
AVGComp
sets the instrument to stop acquisition after the number of
waveforms speci
fi
ed by the ACQuire:NUMAVg command have been
acquired and averaged.
Examples
ACQUIRE:STOPAFTER:CONDITION HISTHits
sets the instrument to stop
acquisition after it acquires the speci
fi
ed number of valid hits in the histogram
region.
ACQUIRE:STOPAFTER:CONDITION?
might return
ACQuire:STOPAfter:CONDition MASK3HITS
, indicating that the acquisition
will be stopped after the hits in mask 3 are equal to or greater than the speci
fi
ed
number of hits.
2-62
DSA8300 Programmer Manual
Содержание DSA8300
Страница 1: ...xx DSA8300 Digital Serial Analyzer ZZZ Programmer Manual P077057002 077 0570 02 ...
Страница 2: ......
Страница 3: ...DSA8300 Digital Serial Analyzer ZZZ Programmer Manual xx Revision B www tektronix com 077 0570 02 ...
Страница 8: ...Preface iv DSA8300 Programmer Manual ...
Страница 9: ...Getting Started ...
Страница 10: ......
Страница 15: ...Syntax and Commands ...
Страница 16: ......
Страница 404: ...Commands Listed in Alphabetical Order 2 388 DSA8300 Programmer Manual ...
Страница 405: ...Status and Events ...
Страница 406: ......
Страница 426: ...Status and Events 3 20 DSA8300 Programmer Manual ...
Страница 427: ...Examples ...
Страница 428: ......
Страница 431: ...Appendices ...
Страница 432: ......
Страница 433: ...Appendix A Character Set DSA8300 Programmer Manual A 1 ...
Страница 434: ...Appendix A Character Set A 2 DSA8300 Programmer Manual ...
Страница 446: ...Appendix C Factory Default Setup Values C 8 DSA8300 Programmer Manual ...