Commands Listed in Alphabetical Order
Arguments
C1CLKRec
sets the trigger source to the clock recovery output of the module
in the upper left slot. This trigger source requires either an optical sampling
head that supports clock recovery or the 80A05 clock recovery module.
C3CLKRec
sets the trigger source to the clock recovery output of the module
in the upper right slot. This trigger source requires either an optical sampling
head that supports clock recovery or the 80A05 clock recovery module.
CLKPre
sets the instrument to trigger on an external signal connected to the
CLOCK INPUT/PRESCALE TRIGGER input connector on the instrument
front panel. The input coupling is AC. Triggering is guaranteed for clocks up
to 15 GHz (and will typically work up to 20 GHz). This argument performs
the same function as EXTPrescaler. The query form of TRIGger:SOUrce will
always return EXTPrescaler, regardless if the source was set with the CLKPre
or EXTPrescaler argument.
EXTDirect
sets the instrument to trigger on an external signal connected to
the TRIGGER DIRECT INPUT connector on the instrument front panel. The
input coupling mode is DC and triggering is usable with signals up to at least
3.0 GHz. Use this trigger source for low repetition rate signals that cannot be
triggered using the Clock Input/Prescale Trigger front panel connector.
EXTPrescaler
sets the instrument to trigger on an external signal connected
to the CLOCK INPUT/PRESCALE TRIGGER input connector on the
instrument front panel. The input coupling is AC. Triggering is guaranteed for
clocks up to 15 GHz (and will typically work up to 20 GHz). The query form
of TRIGger:SOUrce will always return EXTPrescaler, regardless if the source
was set with the CLKPre or EXTPrescaler argument.
FREerun
sets the instrument to disable synchronous triggered acquisition;
the instrument randomly triggers to acquire data.
INTClk
sets the instrument to trigger on an internal clock signal. The
INTClk trigger source is a synonym for TDR and is provided for backwards
compatibility with legacy 8000 series sampling oscilloscopes. Setting the
Trigger Source to INTClk and then sending the TRIGGER:SOURCE? returns
the "TDR" synonym.
TDR
sets the instrument to trigger on an internal clock signal. Use the
TRIGger:INTRate command to select any rate between 25 kHz and 300 kHz.
Use this type of trigger source with TDR measurements to synchronize the
generation of TDR step pulses with the subsequent acquisition. The Internal
Clock Out connector on the instrument front panel supplies a sample of the
internal clock.
Examples
TRIGGER:SOURCE EXTDIRECT
sets the trigger source to use the Trigger Direct
Input front panel connector.
2-354
DSA8300 Programmer Manual
Содержание DSA8300
Страница 1: ...xx DSA8300 Digital Serial Analyzer ZZZ Programmer Manual P077057002 077 0570 02 ...
Страница 2: ......
Страница 3: ...DSA8300 Digital Serial Analyzer ZZZ Programmer Manual xx Revision B www tektronix com 077 0570 02 ...
Страница 8: ...Preface iv DSA8300 Programmer Manual ...
Страница 9: ...Getting Started ...
Страница 10: ......
Страница 15: ...Syntax and Commands ...
Страница 16: ......
Страница 404: ...Commands Listed in Alphabetical Order 2 388 DSA8300 Programmer Manual ...
Страница 405: ...Status and Events ...
Страница 406: ......
Страница 426: ...Status and Events 3 20 DSA8300 Programmer Manual ...
Страница 427: ...Examples ...
Страница 428: ......
Страница 431: ...Appendices ...
Страница 432: ......
Страница 433: ...Appendix A Character Set DSA8300 Programmer Manual A 1 ...
Страница 434: ...Appendix A Character Set A 2 DSA8300 Programmer Manual ...
Страница 446: ...Appendix C Factory Default Setup Values C 8 DSA8300 Programmer Manual ...