
Commands Listed in Alphabetical Order
Syntax
MEASUrement:MEAS<x>:REFLevel<x>:METHod {RELative | ABSolute
| HIDelta | LODelta | AOPt}
MEASUrement:MEAS<x>:REFLevel<x>:METHod?
Related Commands
MEASUrement:MEAS<x>:REFLevel<x>:ABSolute:HIGH
MEASUrement:MEAS<x>:REFLevel<x>:ABSolute:LOW
MEASUrement:MEAS<x>:REFLevel<x>:ABSolute:MID
MEASUrement:MEAS<x>:REFLevel<x>:RELative:HIGH
MEASUrement:MEAS<x>:REFLevel<x>:RELative:LOW
MEASUrement:MEAS<x>:REFLevel<x>:RELative:MID
Arguments
RELative
calculates the reference levels as a percentage of the High/Low
amplitude (High amplitude minus the Low amplitude). The default values
are 90% for the high reference level, 10% for the low reference level, and
50% for the mid reference levels. You can set other percentages using the
MEASUrement:MEAS<x>:REFLevel:RELative commands.
ABSolute
uses reference levels set explicitly in absolute user units with
the MEASUrement:MEAS<x>:REFLevel:ABSolute commands (see related
commands above). This method is useful when precise values are required
(for example, when you are designing to published interface speci
fi
cations
such as RS-232-C). The default values are 0 V for the high reference level,
the low reference level, and the mid reference levels.
HIDelta
sets the reference levels as absolute amplitude delta values from
the High level.
LODelta
sets the reference levels as absolute amplitude delta values from
the Low level.
AOPt
sets the reference level to best measure Optical Modulation Amplitude
(OMA) on Pulse waveforms, and is the default selection when taking OMA
measurements on Pulse or NRZ waveforms. OMA measurements are
described in the User Online Help, which is accessible from the instrument
Help menu.
Examples
MEASUREMENT:MEAS1:REFLEVEL1:METHOD RELATIVE
sets the method of
calculating the reference levels to relative for Measurement 1; the default values
are 90% for the high reference level, 50% for the mid reference level, and 10% for
the low reference level.
MEASUREMENT:MEAS8:REFLEVEL1:METHOD?
might return
MEASUREMENT:MEAS8:REFLEVEL1:METHOD ABSOLUTE
, indicating
the reference levels used are set to absolute values in user units.
2-242
DSA8300 Programmer Manual
Содержание DSA8300
Страница 1: ...xx DSA8300 Digital Serial Analyzer ZZZ Programmer Manual P077057002 077 0570 02 ...
Страница 2: ......
Страница 3: ...DSA8300 Digital Serial Analyzer ZZZ Programmer Manual xx Revision B www tektronix com 077 0570 02 ...
Страница 8: ...Preface iv DSA8300 Programmer Manual ...
Страница 9: ...Getting Started ...
Страница 10: ......
Страница 15: ...Syntax and Commands ...
Страница 16: ......
Страница 404: ...Commands Listed in Alphabetical Order 2 388 DSA8300 Programmer Manual ...
Страница 405: ...Status and Events ...
Страница 406: ......
Страница 426: ...Status and Events 3 20 DSA8300 Programmer Manual ...
Страница 427: ...Examples ...
Страница 428: ......
Страница 431: ...Appendices ...
Страница 432: ......
Страница 433: ...Appendix A Character Set DSA8300 Programmer Manual A 1 ...
Страница 434: ...Appendix A Character Set A 2 DSA8300 Programmer Manual ...
Страница 446: ...Appendix C Factory Default Setup Values C 8 DSA8300 Programmer Manual ...