Commands Listed in Alphabetical Order
MEASUrement:MEAS<x>:SOUrce<x>:GATE<x>:PCTPos
This command sets or queries the position (in percentage of the waveform) of
the source gates for the measurement. The measurement, source, and gate are
de
fi
ned as follows:
MEAS<x> where x can be 1 through 8 (a maximum of 8 measurement slots)
SOURCE<x> where x can be 1 or 2 (a maximum of 2 source waveforms)
GATE<x> where x can be 1 or 2 (one pair of gates per source waveform)
Gates are generic, nondirectional markers that segment the area of waveform
data over which measurements are performed. There is no distinction in gates of
stop or start, so they can cross.
You use this command to set the position of a gate as a percentage of the
waveform. This is equivalent to selecting the Region dialog in the Measurement
Setup dialog box and entering the position as a percentage of the waveform in the
G1 or G2 control in the Gates section.
Note that a user cannot directly set the position of a gate in the Measurement Setup
dialog box using waveform units. Instead, the user sets the percentage of the
waveform at which the gate is positioned, and then the position in waveform units
is displayed in the readout boxes adjacent to the G1 and G2 controls. However,
you can directly set the gate position in waveform units with the GPIB command,
MEASUrement:MEAS<x>:SOURCE<x>:GATE<x>:POS.
NOTE.
The SOURCE<x> argument speci
fi
es the source, which can be source 1
or source 2. To set the actual waveform (channel, reference, or math) that is
assigned to source, you must use the MEASUrement:MEAS<x>:SOURCE<x>
WFM command. Two sources are provided because some measurements (gain,
phase) require two sources.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:SOUrce<x>:GATE<x>:PCTPos <NR2>
MEASUrement:MEAS<x>:SOUrce<x>:GATE<x>:PCTPos?
Related Commands
MEASUrement:MEAS<x>:GATing:STATE
Arguments
NR2
is the position of the speci
fi
ed gate in percentage.
Examples
MEASUREMENT:MEAS4:SOURCE1:GATE1:PCTPOS 25
sets Gate 1 to 25% on the
Source 1 waveform used for Measurement 4.
DSA8300 Programmer Manual
2-249
Содержание DSA8300
Страница 1: ...xx DSA8300 Digital Serial Analyzer ZZZ Programmer Manual P077057002 077 0570 02 ...
Страница 2: ......
Страница 3: ...DSA8300 Digital Serial Analyzer ZZZ Programmer Manual xx Revision B www tektronix com 077 0570 02 ...
Страница 8: ...Preface iv DSA8300 Programmer Manual ...
Страница 9: ...Getting Started ...
Страница 10: ......
Страница 15: ...Syntax and Commands ...
Страница 16: ......
Страница 404: ...Commands Listed in Alphabetical Order 2 388 DSA8300 Programmer Manual ...
Страница 405: ...Status and Events ...
Страница 406: ......
Страница 426: ...Status and Events 3 20 DSA8300 Programmer Manual ...
Страница 427: ...Examples ...
Страница 428: ......
Страница 431: ...Appendices ...
Страница 432: ......
Страница 433: ...Appendix A Character Set DSA8300 Programmer Manual A 1 ...
Страница 434: ...Appendix A Character Set A 2 DSA8300 Programmer Manual ...
Страница 446: ...Appendix C Factory Default Setup Values C 8 DSA8300 Programmer Manual ...