Check analog bandwidth
This test checks the bandwidth at 50 Ω termination for each channel. The typical
bandwidth at 1 M Ω termination is checked on the MSO/DPO5200/B and MSO/
DPO5100/B series products as a functional check.
1.
Connect the output of the leveled sine wave generator to the oscilloscope
channel 1 input as shown in the following illustration.
WARNING.
The generator is capable of providing dangerous voltages. Be
sure to set the generator to off or 0 volts before connecting, disconnecting,
and/or moving the test hookup during the performance of this procedure.
2.
Push the front-panel
Default Setup
button.
3. Set the impedance to 50 Ω as follows:
a.
Set the calibrator to 50 Ω output impedance, and to generate a sine wave.
b.
Push the front-panel oscilloscope Vertical
V
menu button.
c.
Under
Termination
, click
50 Ω
.
4.
Set the Vertical
Scale
to
1 mV
.
5. Set the Acquisition mode to Sample as follows:
a.
Push the front-panel
Acquire
button.
b.
In the Acquisition window, select the
Acquisition
tab at the left.
c.
Click
Sample
Acquisition mode.
6.
Adjust the signal source to 8 vertical divisions at the selected vertical scale
with a set frequency of 50 kHz. For example, at 5 mV/div, use a ≥40 mV
p-p
signal, and at 2 mV/div, use a ≥16 mV
p-p
signal.
NOTE.
At some V/div settings, the generator may not provide 8 vertical
divisions of signal. Use the maximum output setting of the generator to
obtain as many vertical divisions of signal as possible.
7.
Set the Horizontal
Scale
to
1 ms/division
.
Performance verification (MSO/DPO5000/B series)
326
MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series