Ch
Scale
setting
Termination
setting
Position
setting
(Divs)
Offset
setting
Generat
or
setting
Measurement
mean
Difference of
measurement
means
Accuracy limits
Ch4
140 mV
0 V
0
–1 V
–580 mV
+823.2 mV to
+856.8 mV
–1.42 V
0
0 V
420 V
+823.2 mV to
+856.8 mV
–420 mV
0
+1 V
1.42 V
+823.2 mV to
+856.8 mV
580 mV
Ch4
200 mV
0 V
0
–0.9 V
–300 mV
+1.176 V to
+1.224 V
–1.50 V
0
0 V
600 mV
+1.176 V to
+1.224 V
–600 mV
0
+0.9 V
1.50 mV
+1.176 V to
+1.224 V
300 V
Ch4
400 mV
0 V
0
–0.3 V
900 mV
+2.352 mV to
+2.448 mV
–1.50 V
0
0 V
1.20 V
+2.352 mV to
+2.448 mV
–1.20 V
0
+0.3 V
1.50 V
+2.352 mV to
+2.448 mV
–900 mV
d. Display the test signal:
■
Display the
Vertical Setup
menu.
■
Use the keypad to set the Termination Voltage, Position, and Offset
vertical settings as listed in the table for the current vertical scale
setting. The baseline level may move off screen.
■
Set the generator so that its Sense value and polarity matches the
generator setting in the table for the vertical scale, position, offset,
and termination settings you have made. The DC test level should
appear at about 3 division on screen. Using the Vertical Position
knob, move the test level to 3 divisions above the center screen.
NOTE.
It’s important to use the Sense reading of the generator instead of
the Output level when adjusting the generator to the specified voltage
setting.
Performance verification (MSO/DPO70000C, MSO/DPO70000DX, and DPO7000C series)
MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series
209