f. Wait:
Signal-path compensation may take five to fifteen minutes to run.
g. Confirm signal-path compensation returns passed status:
Verify that the
word
Pass
appears in the instrument calibration control window.
2. Return to regular service:
Click the
X
(close) button to exit the instrument
calibration control window.
Functional tests
The purpose of these procedures is to confirm that the instrument functions
properly. The only equipment required is a BNC or SMA cable and an adapter or
the instrument probe. If you need to store settings during these procedures, access
the local C: drive and store them in the TekScope > Setups directory.
STOP
. These procedures verify functions; that is, they verify that the instrument
features operate. They do not verify that they operate within limits.
Therefore, when the instructions in the functional tests that follow call for you to
verify that a signal appears on-screen "that is about five divisions in amplitude"
or "has a period of about six horizontal divisions," etc., do NOT interpret the
quantities given as limits. Operation within limits is checked in Performance
Tests. (See
Performance tests (MSO/DPO70000C, MSO/DPO70000DX, and
on page 115.)
STOP
. DO NOT make changes to the front-panel settings that are not called out
in the procedures. Each verification procedure will require you to set the
instrument to certain default settings before verifying functions. If you make
changes to these settings, other than those called out in the procedure, you may
obtain invalid results. In this case, just redo the procedure from step 1.
When you are instructed to press a front-panel or screen button, the button may
already be selected (its label will be highlighted). If this is the case, it is not
necessary to press the button.
Performance verification (MSO/DPO70000C, MSO/DPO70000DX, and DPO7000C series)
MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series
103
Содержание DPO5000 Series
Страница 2: ......
Страница 6: ......
Страница 352: ...Index 340 MSO70000C DX DPO70000C DX DPO7000C MSO5000 B DPO5000 B Series...