Leave the trigger system triggered on the positive slope of the
waveform before proceeding to the next check.
■
Press the Trigger
Source
button to toggle it to
Ch 1
.
4. Confirm that the A trigger system is within sensitivity limits (full bandwidth):
a. Set the Horizontal Scale:
Set the Horizontal
Scale
to 200 ps.
b. Display the test signal:
■
Set the generator frequency to full bandwidth as follows:
Model
Generator frequency
≥ 4 GHz models
4 GHz
DPO7354C
3.5 GHz
DPO7254C
2.5 GHz
DPO7104C
1 GHz
DPO7054C
500 MHz
■
Set the generator amplitude on screen as follows:
Model
Amplitude
≥ 4 GHz models
2 divisions
DPO7354C
2.5 divisions
DPO7254C
6 divisions
DPO7104C
6 divisions
DPO7054C
6 divisions
■
Now fine adjust the generator output until the
Ch 1 Amplitude
readout indicates the amplitude is as follows (readout may fluctuate):
Model
Amplitude
≥ 4 GHz models
200 mV
DPO7354C
250 mV
DPO7254C
600 mV
DPO7104C
600 mV
DPO7054C
600 mV
■
Disconnect the cable at
Ch 1
and reconnect it to
Ch 1
through an
attenuator:
Performance verification (MSO/DPO70000C, MSO/DPO70000DX, and DPO7000C series)
262
MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series