Check aux trigger out
Equipment required
Prerequisites
One precision 50 Ω coaxial cable (Item 4)
One 2X attenuator (Item 33)
(See
on page 116.) Also, the
instrument must have passed the DC Voltage
Measurement Accuracy. (See
on page 176.)
1. Install the test hookup and preset the instrument controls:
Figure 35: Aux Trigger Out test hookup
a. Hook up test-signal source:
Connect
Aux Out
to
Ch 1
through a 50 Ω
precision cable and an adapter. (See Figure 35: Aux Trigger Out test
hookup.)
b. Initialize the instrument:
Press the
Default Setup
button.
c. Modify the initialized front-panel control settings:
■
Press the Vertical
Ch 1
button to toggle it off.
■
Set the Horizontal
Scale
to 200 μs.
■
From the button bar, touch
Horiz/Acq
and select the
Acquisition
tab.
■
Touch
Average
and set the number of averages to
64
.
■
Touch the
X
(close) button.
2. Confirm Aux Out is within limits for logic levels:
a. Display the test signal:
■
Press the Vertical
Ch 1
button to display that channel.
■
< 4 GHz models: Touch
Vertical
, select
Vertical Setup
, and then
touch Termination
50 Ω
.
Performance verification (MSO/DPO70000C, MSO/DPO70000DX, and DPO7000C series)
266
MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series