Check probe compensation or fast edge output
Equipment required
Prerequisites
One BNC to Minigrabber adapter (item 18)
One precision 50 Ω coaxial cable (Item 21)
One DC calibration generator (Item 6)
One adapter (Item 19)
(See
on page 116.) Also, the
instrument must have passed Check Timebase
and Delay Time Accuracy and Reference. (See
Check timebase and delay time accuracy and
reference
on page 242.)
1. Install the test hookup and preset the instrument controls:
a. Hook up test-signal:
Refer to the following figure.
Figure 37: Probe compensation or fast edge output initial test hookup
b. Initialize the instrument:
Press the
Default Setup
button.
c. Modify the initialized front-panel control settings:
■
Set the
Vertical Scale
to
200 mV
/div (≤20 GHz models) or
120 mV
/div (>20 GHz models).
■
Set the Horizontal
Scale
to 200 μs.
■
Press
Push-Set 50%
.
■
Use the Vertical
Position
knob to center the display on screen.
■
From the button bar, touch
Horiz/Acq
and select the
Acquisition
tab.
■
Touch
Average
and set the number of averages to
128
.