Verify the file system.
Equipment required
Prerequisites
<4 GHz models: One precision 50 Ω coaxial
cable (Item 4)
<4 GHz models: One BNC to Minigrabber
adapter (item 18)
≥4 GHz models: One SMA cable (item 21)
≥4 GHz models: One adapter (item 19)
None
1. Initialize the instrument:
Push the front-panel
Default Setup
button.
2. Hook up the signal source:
Connect the probe compensation or fast edge
output to the Ch 1 input as shown in the following.
Figure 6: Setup for the file system test
3. Set up the instrument:
Push the front panel
Autoset
button.
4.
Set the
Vertical Scale
to
200 mV
/div (≤20 GHz models) or
120 mV
/div
(>20 GHz models).
5. Set the time base:
Set the horizontal
Scale
to
1 ms/div
. The time-base
readout is displayed at the bottom of the graticule.
6. Save the settings:
a.
Pull down the
File
menu to select
Save As. . . >Setup. . .
. This displays
the instrument Save As control window.
b.
Note the default location and file name and then click the
Save
button to
save the setup to the default file name and location.
7. Change the settings again:
Set the horizontal
SCALE
to
200μs/div
.
8. Verify the file system works:
a.
Pull down the
File
menu to select
Recall
. This displays the instrument
Recall control window.
b.
Click Recall
What > Setup
.
Performance verification (MSO/DPO70000C, MSO/DPO70000DX, and DPO7000C series)
MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series
111