Calibration Procedure—442 Service
Adjustment Procedure
If the Line Voltage Selector (120 V/240 V) setting is
changed, the line fuse (located in a fuse holder on
the rear panel) must be of the value specified for the
new Selector setting. Refer to the Operating Instruc
tions section of this manual for complete operating
voltage and fuse information.
Unless otherwise specified, the 442 is originally set for
120 V and HI line operation.
2.
Connect the 442 power cord plug to the line voltage
source and press the ON/OFF button on (in). The POWER
ON LED should light.
NOTE
The 442 must have a warmup time of at least 20
minutes before starting the Adjustment Procedure,
and must be adjusted at an ambient temperature
between 20° C and 30° C to ensure that checks in this
procedure will meet specifications listed in Section
1
.
3.
Set the TRIGGERING MODE to AUTO, and
SOURCE to CH 1/CH 2.
4.
Set CPLG to AC (out) and leave it there throughout
the procedure unless instructed otherwise.
5. Set the following 442 controls as listed:
VERT MODE
CH 1
POSITION (CH 1 & CH 2)
Midrange
VOLTS/DIV (CH 1 & CH 2)
2 m (IX)
VOLTS/DIV VAR (CH 1 & CH 2)
Detent (fully cw)
AC-GND-DC (CH 1)
GND
AC-GND-DC (CH 2)
DC
SEC/DIV
.1 m
MAGNIFIER
XI (out)
POSITION (Horizontal)
Midrange
SLOPE
+ (out)
HOLD OFF
Fully ccw
LEVEL
Fully cw
INTENSITY
Midrange
FOCUS
Midrange
6.
A baseline trace should be visible within the
graticule area. Adjust the INTENSITY and FOCUS con
trols for a low-intensity, well-defined trace.
NOTE
The ASTIG control also affects trace definition, and
if necessary, should be adjusted after a signal has
been applied, for best overall definition of a dis
played signal (use a small-bit screwdriver for this
adjustment).
The baseline trace should be parallel with the horizon
tal graticule lines. If it is not, adjust TRACE ROT (Trace
Rotation) with a small-bit screwdriver until the trace is
aligned.
4-16
REV A MAY 1980
Содержание 442
Страница 9: ...442 Service 442 Oscilloscope viii REV A MAY 1980 ...
Страница 113: ...Figure 7 4 442 Oscilloscope block diagram REV A JUN 1980 2 3 7 4 2 7 A ...
Страница 117: ...A8 VERTICAL BOARD Figure 7 6 Vertical circuit board location ...
Страница 118: ......
Страница 120: ......
Страница 122: ......
Страница 123: ...Figure 7 8 interface circuit board location ...
Страница 124: ......
Страница 126: ......
Страница 130: ...A ll TRIGGER SWITCH BOARD 2 3 7 4 3 6 A Figure 7 10 Trigger Switch circuit board location ...
Страница 131: ...2 3 7 4 S 4 R ev jun VRSO V E R T IC A L OUT PUT CRT t PROBE ADJ 3 ...
Страница 133: ......
Страница 137: ......
Страница 139: ......
Страница 140: ...4 4 Z ...
Страница 142: ......
Страница 145: ...A14 HORIZONTAL OUTPUT BOARD 2 3 7 4 4 0 A Figure 7 14 Horizontal Output circuit board locations ...
Страница 148: ...R EV K JUN 9SO ...
Страница 153: ... 442 ...
Страница 154: ...442 REV A DEC 1978 ...
Страница 164: ......