CALIBRATION TEST EQUIPMENT REPLACEMENT
Calibration Test Equipment Chart
This chart compares TM 500 product performance to that of older Tektronix equipment. Only those
characteristics where significant specification differences occur, are listed. In some cases the new instrument
may not be a total functional replacement. Additional support instrumentation may be needed or a change in
calibration procedure may be necessary.
Comparison of Main Characteristics
DM 501 replaces 7D13
PG 501 replaces 107
PG 501 - Risetime less than
107 - Risetime less than
3.5 ns into 50 O.
3.0 ns into 50 fi.
108
PG 501 - 5 V output pulse;
108 - 10 V output pulse
3.5 ns Risetime
1 ns Risetime
PG 502 replaces 107
108
PG 502 - 5 V output
108 - 10 V output
111
PG 502 - Risetime less than
111 - Risetime 0.5 ns; 30
1 ns; 10 ns
to 250 ns
Pretrigger pulse
Pretrigger pulse
delay
delay
PG 508 replaces 114
Performance of replacement equipment is the same or
115
better than equipment being replaced.
2101
PG 506 replaces 106
PG 506 - Positive-going
106 - Positive and Negative-
trigger output sig-
going trigger output
nal at least 1 V;
signal, 50 ns and 1 V;
High Amplitude out-
High Amplitude output,
put, 60 V.
100 V.
067-0502-01
PG 506 - Does not have
0502-01 - Comparator output
chopped feature.
can be alternately
chopped to a refer-
ence voltage.
SG 503 replaces 190,
190A, 190B
SG 503 - Amplitude range
190B - Amplitude range 40 mV
5 mV to 5.5 V p-p.
to 10 V p-p.
191
067-0532-01
SG 503 - Frequency range
0532-01 - Frequency range
250 kHz to 250 MHz.
65 MHz to 500 MHz.
SG 504 replaces
067-0532-01
SG 504 - Frequency range
0532-01 - Frequency range
245 MHz to 1050 MHz.
65 MHz to 500 MHz.
067-0650-00
TG 501 replaces 180,
180A
TG 501 - Trigger output-
180A - Trigger pulses 1, 10,
slaved to marker
100 Hz; 1, 10, and
output from 5 sec
100 kHz. Multiple
through 100 ns. One
time-marks can be
time-mark can be
generated simultan-
generated at a time.
eously.
181
181 - Multiple time-marks
184
TG 501 - Trigger output-
184 - Separate trigger
slaved to market
pulses of 1 and 0.1
output from 5 sec
sec; 10, 1, and 0.1
through 100 ns. One
ms; 10 and 1 //s.
time-mark can be
generated at a time.
2901
TG 501 - Trigger output-
2901 - Separate trigger
slaved to marker
pulses, from 5 sec
output from 5 sec
to 0.1
n
s. Multiple
through 100 ns.
time-marks can be
One time-mark can
generated simultan-
be generated at
eously.
a time.
NOTE: All TM 500 generator outputs are short-proof. All TM 500 plug-in instruments require TM 500-Series Power Module.
R E V B, J U N 1978
Содержание 442
Страница 9: ...442 Service 442 Oscilloscope viii REV A MAY 1980 ...
Страница 113: ...Figure 7 4 442 Oscilloscope block diagram REV A JUN 1980 2 3 7 4 2 7 A ...
Страница 117: ...A8 VERTICAL BOARD Figure 7 6 Vertical circuit board location ...
Страница 118: ......
Страница 120: ......
Страница 122: ......
Страница 123: ...Figure 7 8 interface circuit board location ...
Страница 124: ......
Страница 126: ......
Страница 130: ...A ll TRIGGER SWITCH BOARD 2 3 7 4 3 6 A Figure 7 10 Trigger Switch circuit board location ...
Страница 131: ...2 3 7 4 S 4 R ev jun VRSO V E R T IC A L OUT PUT CRT t PROBE ADJ 3 ...
Страница 133: ......
Страница 137: ......
Страница 139: ......
Страница 140: ...4 4 Z ...
Страница 142: ......
Страница 145: ...A14 HORIZONTAL OUTPUT BOARD 2 3 7 4 4 0 A Figure 7 14 Horizontal Output circuit board locations ...
Страница 148: ...R EV K JUN 9SO ...
Страница 153: ... 442 ...
Страница 154: ...442 REV A DEC 1978 ...
Страница 164: ......