B. VERTICAL AMPLIFIER
Calibration Procedure—442 Service
Adjustment Procedure
Equipment Required
Amplitude Calibrator (Item 1)
Square-Wave Generator (Item 4)
50 O BNC Termination (Item 20)
50 fi BNC Cable (Item 12)
Low-Capacitance Alignment Tool (Item 10)
Screwdriver (Item 18)
10X Probe (Item 15)
10X Attenuator (Item 11)
Probe-tlp-to-BNC Adapter (Item 9)
Dual Input Coupler (Optional) (Item 13)
Shield (Item 19)
P R E L IM IN A R Y C O N T R O L S E T T IN G S
Preset Front Panel controls as follows:
NOTE
Do not preset internal controls.
VERT MODE
CH 1
VOLTS/DIV (CH 1 & CH 2)
10 m (IX)
VOLTS/DIV VAR (CH 1 & CH 2)
Fully cw (detent)
AC-GND-DC (CH 1 & CH 2)
GND
POSITION (all)
Midrange
SEC/DIV
.5 m
SEC/DIV VAR
Fully cw (detent)
MAGNIFIER
XI (out)
SOURCE
CH 1/CH 2
TRIGGERING MODE
AUTO
SLOPE
+ (out)
CPLG
AC (out)
LEVEL
Midrange
HOLD OFF
Fully ccw
INTENSITY
Forvisibledisplay
Set all other controls as desired. The 442should display
a baseline trace with the foregoing control settings. Set
the INTENSITY, FOCUS, ASTIG, and POSITION controls
as required to maintain a well-defined medium-intensity
display within the graticule area.
NOTE
Verify that the Calibration Fixture Shield has been
installed as described under Preparation in the
Introduction portion of this Adjustment Procedure.
If only a partial Adjustment procedure is to be
performed, refer to Introduction and Preliminary
Procedure at the beginning of this Adjustment
Procedure to prepare the 442 for adjustment.
P R O C E D U R E
1. Vertical Preamplifier Balance
a. ADJUST— CH 1 DC BAL (R4130, see Figure4-9) for
minimum trace shift while switching Channel 1
VOLTS/DIV between 10 m and 20 m settings.
b. Set VERT MODE to CH 2.
c. ADJUST—CH 2 DC BAL (R4230, see Figure4-9) for
minimum trace shift while switching Channel 2
VOLTS/DIV between 10 m and 20 m settings.
2. Differential Balance (CH 2 Invert Balance)
a. Set both VOLTS/DIV switches to 10 m (IX window)
and set VERT MODE to CH 1. Set Channel 1 POSITION
control to align the trace with the center horizontal
graticule line.
b. Set VERT MODE to CH 2 and set Channel 2
POSITION control to align the trace with the center
horizontal graticule line.
c. Set VERT MODE to DIFF.
d. ADJUST—Either R4265 (SN B010409-up) or R4280
(SN B010100-B010408) for minimum trace shift while
alternately pressing the DIFF and CH 2 Vertical Mode
buttons (see Figure 4-9 for location of adjustments).
REV A MAY 1980
4-19
Содержание 442
Страница 9: ...442 Service 442 Oscilloscope viii REV A MAY 1980 ...
Страница 113: ...Figure 7 4 442 Oscilloscope block diagram REV A JUN 1980 2 3 7 4 2 7 A ...
Страница 117: ...A8 VERTICAL BOARD Figure 7 6 Vertical circuit board location ...
Страница 118: ......
Страница 120: ......
Страница 122: ......
Страница 123: ...Figure 7 8 interface circuit board location ...
Страница 124: ......
Страница 126: ......
Страница 130: ...A ll TRIGGER SWITCH BOARD 2 3 7 4 3 6 A Figure 7 10 Trigger Switch circuit board location ...
Страница 131: ...2 3 7 4 S 4 R ev jun VRSO V E R T IC A L OUT PUT CRT t PROBE ADJ 3 ...
Страница 133: ......
Страница 137: ......
Страница 139: ......
Страница 140: ...4 4 Z ...
Страница 142: ......
Страница 145: ...A14 HORIZONTAL OUTPUT BOARD 2 3 7 4 4 0 A Figure 7 14 Horizontal Output circuit board locations ...
Страница 148: ...R EV K JUN 9SO ...
Страница 153: ... 442 ...
Страница 154: ...442 REV A DEC 1978 ...
Страница 164: ......