Removal and Installation Procedures
Maintenance
6Ć24
c.
Next, grasp both the line cord and the retaining clamp and rotate
them 90 degrees counter-clockwise.
d.
Pull the line cord and clamp away to complete the removal.
4.
Remove line fuse: Find the fuse cap on the rear panel. (See Figure 6-6.)
Now, remove the fuse cap by turning it counter clockwise using a
flat-bladed screwdriver, and remove the line fuse.
5.
Reinstallation: Do in reverse steps 3 and 4 to reinstall the line fuse and
then the line cord.
Front Cover, Rear Cover, Cabinet, Rear EMI Gasket, and
Cabinet Handle and Feet
1.
Assemble equipment and locate modules to be removed:
a.
Have handy a screwdriver with a size T-15 Torx
R
tip (Items 1 and 2).
If removing the EMI gasket at the inside rear of the cabinet, also have
handy a pair of needle-nose pliers (Item 5).
b.
Make sure the oscilloscope front cover is installed; if it is not, install it
by snapping its edges over the trim ring.
c.
Locate the rear cover and cabinet in the locator diagram
Cabinet and
Front-Panel Mounted Modules, Figure 6-2.
2.
Orient the oscilloscope:
a.
Pull out on both of the hubs on the cabinet handle to unlock it for
positioning. While holding the hubs unlocked, rotate the handle
towards the bottom of the oscilloscope.
b.
Set the oscilloscope so its face is down with its front cover on the
work surface and its bottom facing you. Reference Figure 6-7 on
page 6-25 as you do the following steps.
3.
Remove the front feet: If removing a front foot for replacement, use a pair
of slip-jaw pliers to firmly grip the foot, then pull with a turning motion to
remove. Reverse the process to install the replacement foot. If no other
modules are being serviced, skip to the end (step 9) of this procedure.
4.
Remove rear cover: Remove the four T-15 Torx
R
screws securing the
rear cover to the oscilloscope. Lift off the rear cover. If no other modules
are being serviced, skip to the end (step 9) of this procedure.
5.
Remove the rear feet: If removing a rear foot for replacement, use your
fingers to press the foot out from inside the rear cover. Reverse process
to install the replacement foot. If no other modules are being serviced,
skip to the end (step 9) of this procedure.
6.
Remove the cabinet:
a.
Remove the T-15 Torx
R
screw at the left side of the oscilloscope.
b.
Grasp the two handle handle hubs and pull them outward as if to
rotate the handle.
Содержание 410
Страница 9: ...Contents x ...
Страница 13: ...Contents xx ...
Страница 19: ...Safety Summary xx ...
Страница 25: ...Introduction x x ...
Страница 26: ...Specifications ...
Страница 47: ...Operating Information ...
Страница 63: ...Operating Information Operating Information 2Ć16 ...
Страница 64: ...Theory of Operation ...
Страница 68: ...Circuit Description Theory of Operation 3Ć4 ...
Страница 69: ...Performance Verification ...
Страница 81: ...Brief Procedures Performance Verification 4Ć12 ...
Страница 119: ...Performance Tests Performance Verification 4Ć50 ...
Страница 120: ...Adjustment Procedures ...
Страница 147: ...Maintenance ...
Страница 151: ...Maintenance Information Maintenance 6Ć4 ...
Страница 228: ...Troubleshooting TDS 410 TDS 420 TDS 460 Service Manual 6Ć81 J2 Pin 1 Figure 6 38 A02 Display Connector J2 ...
Страница 237: ...Troubleshooting Maintenance 6Ć90 J7 Pin 1 J8 Pin 1 J6 Pin 1 W1 Pin 1 W2 Pin 1 Figure 6 46 A01 Backplane Module ...
Страница 240: ...Options ...
Страница 249: ...Electrical Parts List ...
Страница 251: ...Replaceable Electrical Parts List Electrical Parts List 8Ć2 ...
Страница 252: ...Diagrams ...
Страница 258: ...Diagrams 9Ć6 ...
Страница 259: ...Mechanical Parts List ...
Страница 263: ...Mechanical Parts List 10Ć4 ...
Страница 272: ...TDS 410 TDS 420 TDS 460 Service Manual 10Ć13 1 2 3 4 5 6 7 8 9 10 11 6 3 Figure 10 3 Cables and Routing ...
Страница 275: ...Mechanical Parts List 10Ć16 ...