TDS 410, 420 & TDS 460 Service Manual (Company Confidential Final Draft)
4Ć1
Brief Procedures
The
Self Tests use internal routines to confirm basic functionality and proper
adjustment. No test equipment is required to do these test procedures.
The
Functional Tests utilize the probe-adjust output at the front panel as a
test-signal source for further verifying that the oscilloscope functions properly.
A standard-accessory probe, included with this oscilloscope, is the only
equipment required.
These procedures verify the TDS 400 Digitizing Oscilloscopes.
Besides the
Brief Procedures, the set of procedures that can be used to verify
oscilloscope performance includes the
Performance Tests found later in this
section. You may not need to perform all of these procedures, depending on
what you want to accomplish:
H
To rapidly confirm that this oscilloscope functions and is adjusted proper-
ly, just do the procedures under
Self Tests, which begin on page 4-4.
Advantages: These procedures are quick to do, require no external
equipment or signal sources, and perform extensive functional and accu-
racy testing to provide high confidence that the oscilloscope will perform
properly. Use these procedures as a quick check before making a series
of important measurements.
H
To further check functionality, first do the
Self Tests just mentioned; then
do the procedures under
Functional Tests that begin on page 4-6.
Advantages: These procedures require minimal additional time to per-
form, require no additional equipment other than a standard-accessory
probe, and more completely test the internal hardware of this oscillo-
scope. Use these procedures to quickly determine if the oscilloscope is
suitable for putting into service, such as when it is first received.
H
If more extensive confirmation of performance is desired, do the
Perfor-
mance Tests, beginning on page 4-13, after doing the Functional and Self
Tests just referenced.
Advantages: These procedures add direct checking of warranted specifi-
cations. They require more time to perform and suitable test equipment is
required. (See
Equipment Required on page 4-13.)
If you are not familiar with operating this oscilloscope, read
General Operating
Instructions in Section 3 of this manual. These instructions will acquaint you
with the use of the front-panel controls and the menu system.
General Instructions
Содержание 410
Страница 9: ...Contents x ...
Страница 13: ...Contents xx ...
Страница 19: ...Safety Summary xx ...
Страница 25: ...Introduction x x ...
Страница 26: ...Specifications ...
Страница 47: ...Operating Information ...
Страница 63: ...Operating Information Operating Information 2Ć16 ...
Страница 64: ...Theory of Operation ...
Страница 68: ...Circuit Description Theory of Operation 3Ć4 ...
Страница 69: ...Performance Verification ...
Страница 81: ...Brief Procedures Performance Verification 4Ć12 ...
Страница 119: ...Performance Tests Performance Verification 4Ć50 ...
Страница 120: ...Adjustment Procedures ...
Страница 147: ...Maintenance ...
Страница 151: ...Maintenance Information Maintenance 6Ć4 ...
Страница 228: ...Troubleshooting TDS 410 TDS 420 TDS 460 Service Manual 6Ć81 J2 Pin 1 Figure 6 38 A02 Display Connector J2 ...
Страница 237: ...Troubleshooting Maintenance 6Ć90 J7 Pin 1 J8 Pin 1 J6 Pin 1 W1 Pin 1 W2 Pin 1 Figure 6 46 A01 Backplane Module ...
Страница 240: ...Options ...
Страница 249: ...Electrical Parts List ...
Страница 251: ...Replaceable Electrical Parts List Electrical Parts List 8Ć2 ...
Страница 252: ...Diagrams ...
Страница 258: ...Diagrams 9Ć6 ...
Страница 259: ...Mechanical Parts List ...
Страница 263: ...Mechanical Parts List 10Ć4 ...
Страница 272: ...TDS 410 TDS 420 TDS 460 Service Manual 10Ć13 1 2 3 4 5 6 7 8 9 10 11 6 3 Figure 10 3 Cables and Routing ...
Страница 275: ...Mechanical Parts List 10Ć16 ...